8.3 Software and Functional Checks after Installation/Upgrade procedure
1.
Press the Utilities key on the control console.
2.
In the “Utilities” menu touch the
System Setup
button to invoke the setup desktop on the screen.
3.
On the left side of the screen select
Administration
and then click the
Options
tab.
4.
Verify the correct settings of the
Options
. If necessary, customize
the settings according to the printout.
5.
Click the
System Info
tab.
Figure 8-19 System Setup - Administration - System Info page
1 Ultrasound Application Software version
3 Tune version of Application presets
2 Backup File Name
6.
Check the Application Software version.
7.
Check that the Tune version of the Application presets match the Application Software version.
Note
It is neither required nor advisable to reload a previously stored “Full Backup” after a software upgrade that
was performed by means of the FMI from DVD button! If the Tune version does not match the Application
Software version, a warning message appears whenever booting up the system. In this case, it is essential
to load the proper Application Settings (image presets), adapted for the installed software version. Refer to:
Section 5.14.3.10 "Load Application Settings and Factory Presets" on page 5-44
8.
Perform basic functional checks to ensure system is functioning normally.
9.
Check Service Connectivity; if required, perform InSite Checkout. i.e., Service platform has to be
configured properly; see
Section 7.6.5 "Configuration" on page 7-14
.
Replacement Procedures
Voluson E-Series Service Manual
KTD106657 Revision 2
8-13
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...