5.8 Peripherals
5.8.1 Recording Tools
DVD Recorder
The DVD recorder is controlled from the systems front panel. Operation is completely independent of the
Voluson E-Series system. One must set the appropriate Recorder Type, Video Norm (PAL or NTSC) and
Region Code in accordance with the regional standards.
USB Recorder
The optional USB recorder is controlled via included remote control "Rec" switch. Operation is completely
independent of the Voluson E-Series system. For video recording use an USB stick FAT32 or USB-HDD that
is powered from an external source. The movies recorded can not be replayed on the Voluson E-Series
system! For recommended players see user manual of the USB video recorder.
5.8.2 Printers
Black & White Digital Printer
The B&W Digital Printer receives image data via the USB port. The print command is controlled by the keys
P1, P2, P3 and/or P4 on the Voluson E-Series control console (depending on system configuration).
Color Digital Printer
The Color Digital Printer receives image data via the USB port. The print command is controlled by the keys
P1, P2, P3 and/or P4 on the Voluson E-Series control console (depending on system configuration).
Color Deskjet Printer
A Color Deskjet Printer is used to print out reports and exams, but in some cases also ultrasound images.
Usually it is controlled via Bluetooth Adapter.
Network Color Laser Printer
The Network Color Laser Printer is used to print out reports and exams, but in some cases also ultrasound
images. The printer can either be connected directly to the Voluson E-Series system, or to work within the
network environment (usually the hospital network).
5.8.3 DVD Drive
Caution
Laser radiation: Avoid exposure to the beam Class 3B laser product.
Class 3B laser radiation: When open avoid exposure to the beam.
DVD Drive
The DVD Drive (Writer) is used to backup images and reports. In addition, it is used as the main source of
software upgrades and other service utility operations. It is controlled by the BEP via USB port.
5.8.4 ECG-preamplifier (MAN - optional)
The ECG-preamplifier is used for acquiring an ECG-signal to be displayed with the ultrasound image. This
optional peripheral serves for gaining an ECG-signal to mark the systolic and end diastolic moments in M-
Mode and Doppler evaluations.
The ECG-preamplifier must not be used for ECG-diagnostics. It is not intended for use as a cardiac monitor
and must not be used for an intra-operative application on the heart.
Components and Functions (Theory)
Voluson E-Series Service Manual
KTD106657 Revision 2
5-33
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...