3.13.2.10 WLAN Diagnostic
Running diagnostics is good if you think the adapter might be bad. Diagnostics also tell you if the connection
is behaving properly. Sometimes the system connects by accident to the DVR or another internal device
which uses the TCP/IP protocol. The diagnostics would show if there are errors when performing the full two-
way communication. For example, if the IP address starts with 197 or 169 (loopback addresses), then
something is wrong.
1.
Press the Utilities key on the control console.
2.
On the left side of the screen select
Connectivity
and then click the
Device Setup
tab.
3.
Click the
WLAN Configuration
.
4.
The Wireless Network Configuration tool with available Wireless Networks appear.
5.
Select the
Diagnostics
tab and then click
Run Diagnostics
.
Figure 3-72 Run Diagnostics
3.13.2.11 Repairing the WLAN
Occasionally you may need to repair a WLAN that has lost its connection to the Voluson E-Series.
1.
Press the Utilities key on the control console.
2.
On the left side of the screen select
Connectivity
and then click the
Device Setup
tab.
3.
Click the
WLAN Configuration
.
4.
The Wireless Network Configuration tool with available Wireless Networks appear.
5.
Select the
Diagnostics
tab and then click
Repair
.
Note
DO NOT cancel the Repair operation after you have selected to repair the Wireless LAN connection.
Setup Instructions
Voluson E-Series Service Manual
KTD106657 Revision 2
3-75
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...