Among other significant features of the Voluson E-Series ultrasound system are the following:
•
Integrated FrontEnd (uses advanced ASIC and FGPA technologies)
•
Open connectivity using USB ports
•
Bluetooth - wireless connectivity
•
high performance 19" LCD color monitor
•
Low profile, backlit Keyboard
Voluson E-Series has a digital beam forming system (incorporated in the FrontEnd) which can handle up to
256 element probes by use of multiplexing.
Signal flow from the Probe Connector Panel, to the FrontEnd (FE) Electronics, to the BackEnd Processor
(BEP), and finally is displayed on the LCD monitor and peripherals.
Voluson E-Series internal electronics are divided into three:
•
FrontEnd (FE)
•
BackEnd Processor (BEP)
•
Power Supply Unit
Interconnecting signals from FrontEnd, BackEnd, keyboard, monitor, and power distribution sub-systems are
routed internally.
Major System Components
•
•
BackEnd processor:
•
Control Console (User interface); System I/O with hard keys, Touch Panel and EL-Display:
•
Monitor:
•
•
•
Power supply and Isolation transformer for the peripherals,
•
System mechanical chassis: trolley to keep all major components,
Components and Functions (Theory)
Voluson E-Series Service Manual
KTD106657 Revision 2
5-3
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...