4.3.10 Probe/Connectors Usage
4.3.10.1 Connecting a probe
1.
Place the probe's carrying case on a stable surface and open the case.
2.
Carefully remove the probe and unwrap the probe cable.
3.
DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable
damage.
4.
Turn the connector locking handle counterclockwise.
5.
Align the connector with the probe port and carefully push into place.
6.
Turn the connector locking handle clockwise to secure the probe connector.
7.
Open the side door, lay the cable into the intended cable holders and close the door. So it is free to
move, but not resting on the floor.
4.3.10.2 Activating a probe
1.
Press the Probe key to activate the “Probe Select” menu.
2.
Select the appropriate probe by touching the corresponding key.
3.
Upon selection of an “Application”, the programmed user presets appear.
4.
Touching a “Setting” key causes loading of the preset.
The probe is initialized, the Touch Panel shows the main menu (2D mode) and the ultrasound image
appears on the monitor in write mode (real time display).
4.3.10.3 Deactivating a probe
When deactivating the probe, the probe is automatically placed in standby mode (read mode).
1.
Press the Freeze key.
2.
Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual of Voluson E-
Series for complete cleaning instructions.)
3.
Carefully slide the probe around the side of the keyboard, toward the probe holder. Ensure that the
probe is placed gently in the probe holder.
4.3.10.4 Disconnecting a probe
Prior to disconnect a probe freeze the image. It is not necessary to switch the system off.
1.
Press the Freeze key.
2.
Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual of Voluson E-
Series for complete cleaning instructions.)
3.
Carefully slide the probe around the side of the keyboard, toward the probe holder. Ensure that the
probe is placed gently in the probe holder.
Functional Checks
4-26
Voluson E-Series Service Manual
KTD106657 Revision 2
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...