4.4.2 Load Small Backup (Scan Settings)
The loading procedure overwrites existing image/scan settings on the local hard drive.
Make sure to insert the correct System DVD. Additionally you can load the image settings from “D:
\UserSettings\FactoryDefault\V830”.
4.4.2.1 Preparations
1.
Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.
Press the Utilities key on the control console.
3.
In the “Utilities” menu touch the
System Setup
button to invoke the setup desktop on the screen.
4.
On the left side of the screen select
Backup
and then click the
System Configuration
tab.
5.
Click the
Load
button (1) of the "Small Backup (Scan Settings)" group.
Figure 4-23 Load window
6.
Choose the media (2) and click the
Load
button (3).
Note
If it is desired to load settings from media Internal HDD, click on the Change folder button, browse for the
folder on “D:\usersettings\FactoryDefault\V830” and then click the Load button.
Two settings are available which slightly differ:
-
General = for rest of the world
-
USA = for United States (e.g., image orientation UP/DOWN inverted at transvaginal probes)
7.
Select the appropriate file and click
OK
.
8.
Select the desired loading procedure:
-
"Load "Complete Backup"" on page 4-32
-
"Load only parts of the "Complete Backup"" on page 4-33
Functional Checks
Voluson E-Series Service Manual
KTD106657 Revision 2
4-31
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...