Operating Instructions

Proline Promass Q 500

PROFIBUS PA

Coriolis flowmeter

Keep cover

h

tig

t w

hile

S

p

a

n

n

u

n

g

öf

fn

en

N

ic

h

t

u

n

te

r

ci

rc

uit

sar

ealive

Kee

pc

ove

rt

ig

ht

BA01556D/06/EN/02.16

71342163
Valid as of version

01.00.zz (Device firmware)

Содержание Proline Promass Q 500

Страница 1: ...Proline Promass Q 500 PROFIBUS PA Coriolis flowmeter Keep cove r h ti g t w h i l e S p a n nu ng f f n e n N ic h t u n t e r c i r c uitsarealive K e e p c o v e r t i g h t BA01556D 06 EN 02 16 713...

Страница 2: ...uals or the facility read the Basic safety instructions section carefully as well as all other safety instructions in the document that are specific to working procedures The manufacturer reserves the...

Страница 3: ...2 6 1 Installation conditions 22 6 1 1 Mounting position 22 6 1 2 Requirements from environment and process 24 6 1 3 Special mounting instructions 25 6 2 Mounting the measuring device 28 6 2 1 Require...

Страница 4: ...g device 98 10 3 Connecting via FieldCare 98 10 4 Configuring the device address via software 98 10 4 1 PROFIBUS network 98 10 5 Setting the operating language 98 10 6 Configuring the measuring device...

Страница 5: ...11 Resetting the measuring device 234 12 11 1 Function scope of the Device reset parameter 235 12 12 Device information 235 12 13 Firmware history 237 13 Maintenance 238 13 1 Maintenance tasks 238 13...

Страница 6: ...to a dangerous situation Failure to avoid this situation can result in minor or medium injury NOTICE NOTE This symbol contains information on procedures and other facts which do not result in persona...

Страница 7: ...aning Permitted Procedures processes or actions that are permitted Preferred Procedures processes or actions that are preferred Forbidden Procedures processes or actions that are forbidden Tip Indicat...

Страница 8: ...ing Instructions are aimed at specialists with responsibility for installing the measuring device Incoming acceptance and product identification Storage and transport Installation Transmitter Brief Op...

Страница 9: ...Proline Promass Q 500 PROFIBUS PA Document information Endress Hauser 9 1 4 Registered trademarks PROFIBUS Registered trademark of the PROFIBUS User Organization Karlsruhe Germany...

Страница 10: ...use in hazardous areas in hygienic applications or where there is an increased risk due to process pressure are labeled accordingly on the nameplate To ensure that the measuring device remains in pro...

Страница 11: ...risk of electric shock gloves must be worn 2 4 Operational safety Risk of injury Operate the device in proper technical condition and fail safe condition only The operator is responsible for interfere...

Страница 12: ...dware write protection is disabled when the device is delivered 146 2 7 2 Protecting access via a password Different passwords are available to protect write access to the device parameters or access...

Страница 13: ...ss parameter This does not affect cyclic measured value transmission to the higher order system which is always guaranteed For detailed information see the Description of Device Parameters document pe...

Страница 14: ...rder code for Integrated ISEM electronics option A Sensor For use in applications not required to meet special requirements due to ambient or operating conditions As the electronics are located in the...

Страница 15: ...sor operation in underground installations Permanent sensor immersion in water N ic h t u n t e r c i r c uitsarealive 4 7 Nich t un ter Sp an nu ng ffnen Do no t op en wh en en ergize d Ne pa s ou vr...

Страница 16: ...Order code Ser no Ext ord cd i i Date A0029317 Do the nameplate data match the ordering information on the delivery note A0028673 A0029318 Is the CD ROM with the Technical Documentation depends on de...

Страница 17: ...Enter the serial number from the nameplate www endress com deviceviewer The Endress Hauser Operations App Enter the serial number from the nameplate or scan the 2 D matrix code QR code on the namepla...

Страница 18: ...tion data available inputs and outputs 9 2 D matrix code 10 Manufacturing date year month 11 Document number of safety related supplementary documentation 12 Space for approvals and certificates e g C...

Страница 19: ...for explosion protection Pressure Equipment Directive and degree of protection 8 Flow direction 9 Manufacturing date year month 10 2 D matrix code 11 Document number of safety related supplementary do...

Страница 20: ...ymbol Meaning WARNING This symbol alerts you to a dangerous situation Failure to avoid this situation can result in serious or fatal injury Reference to documentation Refers to the corresponding devic...

Страница 21: ...and dust free place Do not store outdoors Storage temperature 50 to 80 C 58 to 176 F 5 2 Transporting the product Transport the measuring device to the measuring point in the original packaging A00292...

Страница 22: ...Measuring device secondary packaging polymer stretch film that conforms to EC Directive 2002 95 EC RoHS Packaging Wood crate treated in accordance with ISPM 15 standard which is confirmed by the affix...

Страница 23: ...ller cross section than the nominal diameter prevent the sensor running empty while measurement is in progress 1 2 3 4 5 A0028773 6 Installation in a down pipe e g for batching applications 1 Supply t...

Страница 24: ...r inhomogeneous media 1 2 A0028774 7 Orientation of sensor with curved measuring tube 1 Avoid this orientation for fluids with entrained solids Risk of solids accumulating 2 Avoid this orientation for...

Страница 25: ...e the rupture disk When using a rupture disk do not use a heating jacket Make sure that the function and operation of the rupture disk is not impeded through the installation of the device Take precau...

Страница 26: ...t technology Calibration takes place under reference conditions 253 Therefore a zero point adjustment in the field is generally not required Experience shows that zero point adjustment is advisable on...

Страница 27: ...or cable provided by the customer It is recommended to use stainless steel cables or chains If a protective coating is applied it is recommended to use a heat shrink tube to protect the housing paint...

Страница 28: ...6 0 mm For sensor For flanges and other process connections Corresponding mounting tools 6 2 2 Preparing the measuring device 1 Remove all remaining transport packaging 2 Remove any protective covers...

Страница 29: ...ximum ambient temperature If operating outdoors Avoid direct sunlight and exposure to weathering particularly in warm climatic regions LCAUTION Excessive force can damage the housing Avoid excessive m...

Страница 30: ...ews and mount in place 5 Tighten the securing screws 6 2 5 Mounting the transmitter housing Proline 500 LCAUTION Ambient temperature too high Danger of electronics overheating and housing deformation...

Страница 31: ...0 3 94 100 3 94 8 6 0 39 18 0 71 10 0 39 A0029068 12 Engineering unit mm in 1 Drill the holes 2 Insert wall plugs into the drilled holes 3 Screw in the securing screws slightly at first 4 Fit the tran...

Страница 32: ...6 Turning the transmitter housing Proline 500 To provide easier access to the connection compartment or display module the transmitter housing can be turned N i c ht u n t e r a r e f f n e n E ES C 1...

Страница 33: ...stallation check Is the device undamaged visual inspection Does the measuring device conform to the measuring point specifications For example Process temperature 258 Process pressure refer to the cha...

Страница 34: ...2 in 7 1 2 Requirements for connecting cable The connecting cables provided by the customer must fulfill the following requirements Electrical safety In accordance with applicable federal national reg...

Страница 35: ...optical cover 85 Loop resistance Power supply line maximum 10 Cable length Maximum 300 m 1000 ft see the following table Cross section Cable length 0 34 mm2 AWG 22 80 m 270 ft 0 50 mm2 AWG 20 120 m 4...

Страница 36: ...ximum 150 m 500 ft see the following table Cross section Cable length Termination 2 x 2 x 0 50 mm2 AWG 22 50 m 165 ft 2 x 2 x 0 50 mm2 AWG 22 A B 0 5 mm2 A B 0 5 mm2 3 x 2 x 0 50 mm2 AWG 22 100 m 330...

Страница 37: ...ded for the purpose inside the connection housing The stripped and twisted lengths of cable shield to the ground terminal must be as short as possible 7 1 3 Terminal assignment Transmitter supply volt...

Страница 38: ...r 1 Mount the sensor and transmitter 2 Connection housing sensor Connect connecting cable 3 Transmitter Connect connecting cable 4 Transmitter Connect signal cable and cable for supply voltage NOTICE...

Страница 39: ...smitter to the same potential equalization Only connect the sensor to a transmitter with the same serial number Ground the connection housing of the sensor via the external screw terminal Terminal ass...

Страница 40: ...ensure tight sealing do not remove the sealing ring from the cable entry 4 Strip the cable and cable ends In the case of stranded cables fit ferrules 5 Connect the protective ground 6 Connect the cab...

Страница 41: ...ousing cover 2 Open the housing cover 3 Push the cable through the cable entry To ensure tight sealing do not remove the sealing ring from the cable entry 4 Strip the cable and cable ends In the case...

Страница 42: ...Endress Hauser Connecting the sensor connection housing via the connector For the device version with the order code for Sensor connection housing Option C Ultra compact hygienic stainless 1 2 A00296...

Страница 43: ...ntry To ensure tight sealing do not remove the sealing ring from the cable entry 5 Strip the cable and cable ends In the case of stranded cables fit ferrules 6 Connect the protective ground 7 Connect...

Страница 44: ...6 10 0 4 22 mm 8 24 mm A0029597 1 Loosen the 4 fixing screws on the housing cover 2 Open the housing cover 3 Fold open the terminal cover 4 Push the cable through the cable entry To ensure tight seal...

Страница 45: ...g torque applied to the fixing screws Risk of damaging the plastic transmitter Tighten the fixing screws as per the tightening torque 2 Nm 1 5 lbf ft 11 Tighten the 4 fixing screws on the housing cove...

Страница 46: ...f damaging the electronic components Connect the sensor and transmitter to the same potential equalization Only connect the sensor to a transmitter with the same serial number Ground the connection ho...

Страница 47: ...remove the sealing ring from the cable entry 4 Strip the cable and cable ends In the case of stranded cables fit ferrules 5 Connect the protective ground 6 Connect the cable in accordance with the te...

Страница 48: ...ver 2 Open the housing cover 3 Push the cable through the cable entry To ensure tight sealing do not remove the sealing ring from the cable entry 4 Strip the cable and cable ends In the case of strand...

Страница 49: ...ensure tight sealing do not remove the sealing ring from the cable entry 4 Strip the cable and cable ends In the case of stranded cables also fit ferrules 5 Connect the protective ground 6 Connect the...

Страница 50: ...curing clamp of the connection compartment cover 2 Unscrew the connection compartment cover 3 Squeeze the tabs of the display module holder together 4 Remove the display module holder N i c ht u n t e...

Страница 51: ...e f f n e n E ES C 10 11 22 mm 24 mm A0029816 10 Connect the cable in accordance with the terminal assignment Signal cable terminal assignment The device specific terminal assignment is documented on...

Страница 52: ...ngineering unit mm in 1 To remove a cable from the terminal use a flat blade screwdriver to push the slot between the two terminal holes 2 while simultaneously pulling the cable end out of the termina...

Страница 53: ...2 PROFIBUS PA segment coupler 3 Cable shield the cable shield must be grounded at both ends to comply with EMC requirements observe cable specifications 4 T box 5 Measuring device 6 Local grounding 7...

Страница 54: ...N 3 Analog display unit observe maximum load 4 Transmitter Pulse frequency output 1 2 3 12345 A0028761 19 Connection example for pulse frequency output passive 1 Automation system with pulse frequency...

Страница 55: ...Transmitter Observe input values 248 Current input 2 1 3 A0028915 22 Connection example for 4 to 20 mA current input 1 Power supply 2 External measuring device for reading in pressure or temperature f...

Страница 56: ...essing P o w e r I O 2 I O 3 I O 3 4 1 2 3 4 5 6 7 8 128 64 32 16 8 4 2 1 PROFIBUS address not used Off On 1 2 3 4 5 3 ES C 2 Po we r op en pres s op en pres s ES C E I O 4 Po we r I O 2 I O 3 1 A0029...

Страница 57: ...are addressing To switch addressing from hardware addressing to software addressing set DIP switch No 4 to Off The device address configured in the Device address parameter 106 takes effect after 10 s...

Страница 58: ...loops down before the cable entry water trap A0029278 6 Insert dummy plugs into unused cable entries 7 8 Post connection check Are cables or the device undamaged visual inspection Do the cables used m...

Страница 59: ...ion options 1 2 3 4 5 E ESC 6 A0029295 1 Local operation via display module 2 Computer with Web browser e g Internet Explorer or with operating tool e g FieldCare DeviceCare AMS Device Manager SIMATIC...

Страница 60: ...tion Diagnostics Access status display Output Operating menu for experts Language Operatation Language Parameter 1 Setup Submenu 1 Submenu n Device tag Advanced setup Enter access code Parameter 1 Par...

Страница 61: ...ce Define the medium Display I O configuration Configure the inputs Configure the outputs Configuring the operational display Define the output conditioning Set the low flow cut off Configure partial...

Страница 62: ...he measurement or the communication interface Sensor Configuration of the measurement Output Configure the pulse frequency switch output Input Configuring the status input Output Configuring of the an...

Страница 63: ...r Example Appears only if a diagnostics event is present for this measured variable Measured values Symbol Meaning Mass flow Volume flow Corrected volume flow Density Reference density Temperature Tot...

Страница 64: ...r menu In the wizard Display symbol for wizard Omission symbol for operating menu levels in between Name of current Submenu Wizard Parameters Examples Display Display For more information about the ic...

Страница 65: ...the Diagnostics menu Expert Appears In the menu next to the Expert selection At the left in the navigation path in the Expert menu Submenus wizards parameters Symbol Meaning Submenu Wizard Parameters...

Страница 66: ...ls are available in the input mask of the numeric and text editor Numeric editor Symbol Meaning 0 9 Selection of numbers from 0 to 9 Inserts decimal separator at the input position Inserts minus sign...

Страница 67: ...ition one position to the left Deletes one character immediately to the left of the input position 8 3 4 Operating elements Key Meaning Minus key In a menu submenu Moves the selection bar upwards in a...

Страница 68: ...and takes you to the next higher level If help text is open closes the help text of the parameter Pressing the key for 2 s returns you to the operational display home position With a Wizard Exits the...

Страница 69: ...Hauser 69 2 Press simultaneously The context menu is closed and the operational display appears Calling up the menu via the context menu 1 Open the context menu 2 Press to navigate to the desired men...

Страница 70: ...1 0098 1 0098 1 XXXX XXX X X 10 50 2800 Hz mA 3 4 5 6 7 8 1 2 Display operat Display operat Setup Main menu English Main menu Format display Display Contrast display Display intervall 1 value max Setu...

Страница 71: ...channel is jumped to Enter the direct access code with the corresponding channel number Example Enter 0914 2 Assign process variable parameter For the direct access codes of the individual parameters...

Страница 72: ...BC 1x 001 FT 10 DEFG HIJK LMNO PQRS TUVW XYZ Aa1 ABC 001 FT 10 DEFG HIJK LMNO PQRS TUVW XYZ Aa1 ABC 001 FT 10 DEFG HIJK ABC A a 1 001 FT 10 DEFG HIJK ABC A a 1 001 FT 10 3456 789 Aa1 012 001 FT 10 345...

Страница 73: ...has been defined 1 1 If an incorrect access code is entered the user obtains the access rights of the Operator user role The user role with which the user is currently logged on is indicated by the A...

Страница 74: ...the integrated Web server the device can be operated and configured via a Web browser and via a service interface CDI RJ45 or via a WLAN interface The structure of the operating menu is the same as fo...

Страница 75: ...192 168 1 212 basic html in the address line of the Web browser A fully functional but simplified version of the operating menu structure starts in the Web browser When installing a new firmware vers...

Страница 76: ...connecting cable Configuring the Internet protocol of the computer The following information refers to the default Ethernet settings of the device IP address of the device 192 168 1 212 factory settin...

Страница 77: ...l Enable WLAN reception on the mobile terminal Establishing a connection from the mobile terminal to the measuring device 1 In the WLAN settings of the mobile terminal Select the measuring device usin...

Страница 78: ...5 Current measured values 6 Operating language 7 User role 8 Access code 9 Login 10 Reset access code 141 If a login page does not appear or if the page is incomplete 161 8 4 4 Logging on 1 Select th...

Страница 79: ...evice Load the configuration from the measuring device XML format save configuration Save the configuration to the measuring device XML format restore configuration Export the event list csv file Expo...

Страница 80: ...isabled Port 80 is locked On The complete functionality of the web server is available JavaScript is used The password is transferred in an encrypted state Any change to the password is also transferr...

Страница 81: ...oupler PROFIBUS DP PA 5 PROFIBUS PA network 6 T box 7 Measuring device Service interface Via service interface CDI RJ45 Proline 500 digital transmitter open press 1 2 3 A0029163 27 Connection via serv...

Страница 82: ...Display operation option G 4 line backlit graphic display touch control WLAN 5 6 1 2 2 1 3 4 3 4 3 4 3 4 A0029165 1 Transmitter with integrated WLAN antenna 2 Transmitter with external WLAN antenna 3...

Страница 83: ...ce interface Preparing the mobile terminal Enable WLAN reception on the mobile terminal Establishing a connection from the mobile terminal to the measuring device 1 In the WLAN settings of the mobile...

Страница 84: ...7 P P P P 4 8 9 10 11 Xxxxxxx Good Status Device tag Xxxxxxx Device name Mass flow 12 34 kg h Volume flow 12 34 m h Mass flow unit Volume flow unit kg h m h Access status tooling Operation Setup Xxxx...

Страница 85: ...presents a convenient comprehensive solution For details see Innovation Brochure IN01047S Source for device description files See information 86 8 5 4 SIMATIC PDM Function scope SIMATIC PDM is a stand...

Страница 86: ...ons FieldCare www endress com Download Area CD ROM contact Endress Hauser DVD contact Endress Hauser DeviceCare www endress com Download Area CD ROM contact Endress Hauser DVD contact Endress Hauser S...

Страница 87: ...asured values If a system is configured with a Profile GSD it is possible to exchange devices made by different manufacturers However it is essential to ensure that the order of the cyclic process val...

Страница 88: ...t data and measured value status information available for cyclic data exchange If the Promass 500 PROFIBUS PA is acyclically configured via an operating program Class 2 master access is directly via...

Страница 89: ...s replaced the measuring device Promass 500 supports the compatibility of the cyclic data with previous models It is not necessary to adjust the engineering parameters of the PROFIBUS network with the...

Страница 90: ...alizer operation mode parameter in the Totalizer function block 0 9 Measuring mode BIDIRECTIONAL 0 24 UNIT TO BUS No Cause Functionality is no longer required as the unit is adopted automatically 0 25...

Страница 91: ...ndividual modules The device master file GSD contains a description of the individual modules input and output data along with their individual properties The modules are permanently assigned to the s...

Страница 92: ...CHANNEL parameter CHANNEL Input variable 32961 Mass flow 33122 Volume flow 33093 Corrected volume flow 901 Target fluid mass flow 1 793 Carrier mass flow 1 32850 Density 33092 Reference density 794 Co...

Страница 93: ...ion pertaining to the totalizer value Three Totalizer blocks are available slot 9 to 11 Selection totalizer value The totalizer value can be specified using the CHANNEL parameter CHANNEL Input variabl...

Страница 94: ...e module combination consists of the SETTOT MODETOT and TOTAL functions SETTOT Control the totalizers via the PROFIBUS master MODETOT Configure the totalizers via the PROFIBUS master TOTAL Transmit to...

Страница 95: ...ividual Analog Output blocks CHANNEL Function block Compensation value 306 AO 1 External pressure 1 307 AO 2 External temperature 1 488 AO 3 External reference density 1 The compensation values must b...

Страница 96: ...ut data of Discrete Input Byte 1 Byte 2 Discrete Status DO module Discrete Output Transmit discrete output values from the PROFIBUS master Class 1 to the measuring device Discrete output values are us...

Страница 97: ...as a modular PROFIBUS slave In contrast to a compact slave a modular PROFIBUS slave has a variable design and consists of several individual modules The GSD file contains a description of the individ...

Страница 98: ...erational display If nothing appears on the local display or a diagnostic message is displayed refer to the section on Diagnostics and troubleshooting 160 10 3 Connecting via FieldCare For FieldCare 8...

Страница 99: ...local display 10 6 Configuring the measuring device The Setup menuwith its guided wizards contains all the parameters needed for standard operation Navigation to the Setup menu XXXX XXX X X 20 50 010...

Страница 100: ...t 1 to n 109 Status input 1 to n 110 Current output 1 to n 110 Pulse frequency switch output 1 to n 113 Relay output 1 to n 122 Display 125 Low flow cut off 128 Partially filled pipe detection 129 Adv...

Страница 101: ...ring point Max 32 characters such as letters numbers or special characters e g Promass300 500PA 10 6 2 Setting the system units In the System units submenu the units of all the measured values can be...

Страница 102: ...w cut off Simulation process variable Unit choose list Country specific l h gal min us Volume unit Select volume unit Unit choose list Country specific l DN 150 6 m gal us Corrected volume flow unit S...

Страница 103: ...value parameter 6052 Maximum value parameter 6108 Minimum value parameter 6109 Carrier pipe temperature parameter 6027 Maximum value parameter 6029 Minimum value parameter 6030 Reference temperature p...

Страница 104: ...izard submenu contains parameters that must be configured in order to select and set the medium Navigation Setup menu Select medium Medium selection Select medium 105 Select gas type 105 Reference sou...

Страница 105: ...of gas at 0 C 32 F 1 to 99999 9999 m s 415 0 m s Temperature coefficient sound velocity The Others option is selected in the Select gas type parameter Enter temperature coefficient for the gas sound v...

Страница 106: ...OFIBUS PA 106 Endress Hauser Navigation Setup menu Communication Communication Device address 106 Parameter overview with brief description Parameter Description User entry Factory setting Device addr...

Страница 107: ...requisite Description Selection User entry Factory setting Channel Select the process variable Mass flow Volume flow Corrected volume flow Target mass flow Carrier mass flow Density Reference density...

Страница 108: ...I O module 1 to n terminal numbers 108 I O module 1 to n information 108 I O module 1 to n type 108 Apply I O configuration 108 Conversion code 108 Parameter overview with brief description Parameter...

Страница 109: ...used 24 25 I O 2 Signal mode The measuring device is not approved for use in the hazardous area with type of protection Ex i Select the signal mode for the current input Passive Active Passive 0 4 mA...

Страница 110: ...number Shows the terminal numbers used by the status input module Not used 24 25 I O 2 Assign status input Select function for the status input Off Reset totalizer 1 Reset totalizer 2 Reset totalizer...

Страница 111: ...mass flow Carrier mass flow Density Reference density Concentration Temperature Carrier pipe temperature Electronic temperature Oscillation frequency 0 Oscillation amplitude 0 Frequency fluctuation 0...

Страница 112: ...current 0 to 22 5 mA 22 5 mA Failure mode One of the following options is selected in the Assign current output parameter 111 Mass flow Volume flow Corrected volume flow Target mass flow Carrier mass...

Страница 113: ...h output Pulse frequency switch output 1 to n Operating mode 113 Parameter overview with brief description Parameter Description Selection Factory setting Operating mode Define the output as a pulse f...

Страница 114: ...g options is selected in the Assign pulse output parameter 114 Mass flow Volume flow Corrected volume flow Target mass flow Carrier mass flow Enter measured value at which a pulse is output Signed flo...

Страница 115: ...value at minimum frequency 117 Measuring value at maximum frequency 118 Failure mode 118 Failure frequency 119 Invert output signal 119 Parameter overview with brief description Parameter Prerequisit...

Страница 116: ...on amplitude 0 Frequency fluctuation 0 Oscillation damping 0 Oscillation damping fluctuation 0 Signal asymmetry Exciter current 0 HBSI Off Minimum frequency value In the Operating mode parameter the F...

Страница 117: ...cillation damping 0 Oscillation damping fluctuation 0 Signal asymmetry Exciter current 0 Enter maximum frequency 0 0 to 10000 0 Hz 10000 0 Hz Measuring value at minimum frequency In the Operating mode...

Страница 118: ...0 Oscillation damping 0 Oscillation damping fluctuation 0 Signal asymmetry Exciter current 0 Enter measured value for maximum frequency Signed floating point number Depends on country and nominal dia...

Страница 119: ...Mass flow Volume flow Corrected volume flow Target mass flow Carrier mass flow Density Reference density Concentration Temperature Carrier pipe temperature Electronic temperature Oscillation frequenc...

Страница 120: ...21 Assign status 121 Switch on value 121 Switch off value 121 Switch on delay 121 Switch off delay 122 Failure mode 122 Invert output signal 122 Parameter overview with brief description Parameter Pre...

Страница 121: ...n the Operating mode parameter The Flow direction check option is selected in the Switch output function parameter Select process variable for flow direction monitoring Off Volume flow Mass flow Corre...

Страница 122: ...utput behavior in alarm condition Actual status Open Closed Open Invert output signal Invert the output signal No Yes No Visibility depends on order options or device settings 10 6 11 Configuring the...

Страница 123: ...matic viscosity Temp compensated dynamic viscosity Temp compensated kinematic viscosity Temperature Oscillation damping Totalizer 1 Totalizer 2 Totalizer 3 Mass flow Assign diagnostic behavior In the...

Страница 124: ...matically through all the parameters that have to be set for configuring the double pulse output Navigation Setup menu Double pulse output Double pulse output Master terminal number 124 Slave terminal...

Страница 125: ...number Depends on country and nominal diameter Pulse width Define time width of the output pulse 0 5 to 2000 ms 0 5 ms Failure mode Define output behavior in alarm condition Actual value No pulses No...

Страница 126: ...amping 0 Oscillation damping fluctuation 0 Signal asymmetry Exciter current 0 Totalizer 1 Totalizer 2 Totalizer 3 Current output 1 Mass flow 0 bargraph value 1 A local display is provided Enter 0 valu...

Страница 127: ...value 3 A selection was made in the Value 3 display parameter Enter 100 value for bar graph display Signed floating point number 0 Value 4 display A local display is provided Select the measured valu...

Страница 128: ...flow cut off Off Mass flow Volume flow Corrected volume flow Mass flow On value low flow cutoff One of the following options is selected in the Assign process variable parameter 128 Mass flow Volume f...

Страница 129: ...e Select process variable for partially filled pipe detection Off Density Reference density Off Low value partial filled pipe detection One of the following options is selected in the Assign process v...

Страница 130: ...operat Setup Main menu English Main menu Medium selection Setup Setup Advanced setup Ent access code Advanced setup Def access code Device tag Display language A0032223 EN The number of submenus can...

Страница 131: ...rence density 131 Fixed reference density 132 Reference temperature 132 Linear expansion coefficient 132 Square expansion coefficient 132 Parameter overview with brief description Parameter Prerequisi...

Страница 132: ...ient The Calculated reference density option is selected in the Corrected volume flow calculation parameter parameter For media with a non linear expansion pattern enter the quadratic medium specific...

Страница 133: ...igh viscosity fluids Navigation Setup menu Advanced setup Sensor adjustment Zero point adjustment Zero point adjustment Zero point adjustment control 133 Progress 133 Parameter overview with brief des...

Страница 134: ...alizer Select the unit for the process variable of the totalizer Unit choose list Country specific kg lb Control Totalizer 1 to n Control totalizer value Totalize Reset hold Preset hold Totalize Total...

Страница 135: ...n Setup menu Advanced setup Display Display Format display 136 Value 1 display 136 0 bargraph value 1 136 100 bargraph value 1 136 Decimal places 1 136 Value 2 display 136 Decimal places 2 136 Value 3...

Страница 136: ...alizer 1 Totalizer 2 Totalizer 3 Current output 1 Mass flow 0 bargraph value 1 A local display is provided Enter 0 value for bar graph display Signed floating point number Country specific 0 kg h 0 lb...

Страница 137: ...es 4 A measured value is specified in the Value 4 display parameter Select the number of decimal places for the display value x x x x xx x xxx x xxxx x xx Display language A local display is provided...

Страница 138: ...o be set for the WLAN configuration Navigation Setup menu Advanced setup WLAN Settings WLAN settings WLAN IP address 138 Security type 138 WLAN passphrase 138 Assign SSID name 139 SSID name 139 Apply...

Страница 139: ...cel Ok Cancel 10 7 6 Configuration management After commissioning you can save the current device configurationor restore the previous device configuration You can do so using the Configuration manage...

Страница 140: ...es the transmitter data of the device Restore The last backup copy of the device configuration is restored from the device memory to the device s integrated HistoROM The backup copy includes the trans...

Страница 141: ...g numbers letters and special characters Using the parameter to reset the access code Navigation Setup menu Advanced setup Administration Reset access code Reset access code Operating time 141 Reset a...

Страница 142: ...nd the device alarm mode and to verify downstream signal chains switching valves or closed control loops Navigation Diagnostics menu Simulation Simulation Assign simulation process variable 143 Proces...

Страница 143: ...put simulation Switch simulation of the status input on and off Off On Off Input signal level In the Status input simulation parameter the On option is selected Select the signal level for the simulat...

Страница 144: ...output on and off Off On Off Switch status 1 to n The On option is selected in the Switch output simulation 1 to n parameter parameter Select status of the relay output for the simulation Open Closed...

Страница 145: ...de again in the Confirm access code parameter 141 to confirm the code The symbol appears in front of all write protected parameters The device automatically locks the write protected parameters again...

Страница 146: ...ose The user specific access code can then be defined again afterwards Via Web browser FieldCare DeviceCare via CDI RJ45 service interface fieldbus For a reset code contact your Endress Hauser service...

Страница 147: ...on In the Locking status parameter the Hardware locked option is displayed 149 In addition on the local display the symbol appears in front of the parameters in the header of the operational display a...

Страница 148: ...on the local display the symbol appears in front of the parameters in the header of the operational display and in the navigation view XXXX XXX X X 20 50 XX XX A0029425 2 Setting the write protection...

Страница 149: ...rs is temporarily locked on account of internal processes running in the device e g data upload download reset etc Once the internal processing has been completed the parameters can be changed once ag...

Страница 150: ...erview with brief description Parameter Prerequisite Description User interface Mass flow Displays the mass flow currently measured Dependency The unit is taken from the Mass flow unit parameter 102 S...

Страница 151: ...ation unit parameter Signed floating point number Target mass flow With the following conditions Order code for Application package option ED Concentration The software options currently enabled are d...

Страница 152: ...er 0 kg Totalizer status 1 to n Displays the current totalizer status Good Uncertain Bad Totalizer status Hex 1 to n In Target mode parameter the Auto option is selected Displays the current status va...

Страница 153: ...easured values Input values Status input 1 to n Status input 1 to n Value status input 153 Parameter overview with brief description Parameter Description User interface Value status input Shows the c...

Страница 154: ...lculated for the current output 3 59 to 22 5 mA Measured current Displays the current value currently measured for the current output 0 to 30 mA Output values for pulse frequency switch output The Pul...

Страница 155: ...n the Operating mode parameter Displays the current switch output status Open Closed Visibility depends on order options or device settings Output values for relay output The Relay output 1 to n subme...

Страница 156: ...Basic settings using theSetup menu 99 Advanced settings using theAdvanced setup submenu 130 11 6 Performing a totalizer reset The totalizers are reset in the Operation submenu Control Totalizer Functi...

Страница 157: ...oROM application package must be enabled in the device order option for the Data logging submenu to appear This contains all the parameters for the measured value history Data logging is also availabl...

Страница 158: ...Endress Hauser Navigation Diagnostics menu Data logging Data logging Assign channel 1 4 159 Logging interval 159 Clear logging data 159 Data logging 159 Logging delay 159 Data logging control 159 Data...

Страница 159: ...for data logging This value defines the time interval between the individual data points in the memory 0 1 to 999 0 s 1 0 s Clear logging data The Extended HistoROM application package is available Cl...

Страница 160: ...d no output signals The connecting cable is not plugged in correctly 1 Check the connection of the electrode cable and correct if necessary 2 Check the connection of the coil current cable and correct...

Страница 161: ...terminating resistor Not connecting to Web server Web server disabled Using the FieldCare or DeviceCare operating tool check whether the Web server of the measuring device is enabled and enable it if...

Страница 162: ...vaScript not enabled JavaScript cannot be enabled 1 Enable JavaScript 2 Enter http XXX XXX X XXX basic html as the IP address Operation with FieldCare or DeviceCare via CDI RJ45 service interface port...

Страница 163: ...4 Po we r I O 2 I O 3 1 A0029689 1 Supply voltage 2 Device status 3 Not used 4 Communication 5 Service interface CDI active 1 Open the housing cover 2 Remove the display module 3 Fold open the termina...

Страница 164: ...r too low 2 Device status Red Error Flashing red Warning 3 Not used 4 Communication White Communication active 5 Service interface CDI Yellow Connection established Flashing yellow Communication activ...

Страница 165: ...gnostics and troubleshooting Endress Hauser 165 LED Color Meaning 1 Communication White Communication active 2 Device status Red Error Flashing red Warning 3 Supply voltage Green Supply voltage is ok...

Страница 166: ...stic event with the highest priority is shown Other diagnostic events that have occurred can be displayed in the Diagnostics menu Via parameter Via submenus 231 Status signals The status signals provi...

Страница 167: ...fault can be identified using the diagnostic information The short text helps you by providing information about the fault In addition the corresponding symbol for the diagnostic behavior is displayed...

Страница 168: ...diagnostic message 1 Press symbol The Diagnostic list submenu opens 2 Select the desired diagnostic event with or and press The message for the remedial measures for the selected diagnostic event open...

Страница 169: ...vice mode e g during a simulation Out of specification The device is operated Outside its technical specification limits e g outside the process temperature range Maintenance required Maintenance is r...

Страница 170: ...be shown in the Diagnostics menu Via parameter Via submenu 231 Diagnostic information The fault can be identified using the diagnostic information The short text helps you by providing information ab...

Страница 171: ...llowing diagnostic behaviors can be assigned Diagnostic behavior Description Alarm The device stops measurement The totalizers assume the defined alarm condition A diagnostic message is generated Warn...

Страница 172: ...ce status are firmly assigned to the particular diagnostic behavior and cannot be changed individually The diagnostic information is grouped as follows Diagnostic information pertaining to the sensor...

Страница 173: ...Coding hex Category NE107 Alarm BAD Process related 0x28 to 0x2B F Failure Invalid process condition Warning UNCERTA IN Process related 0x78 to 0x7B S Out of specification Invalid process condition Lo...

Страница 174: ...n Oscillation damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter cur...

Страница 175: ...on Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation...

Страница 176: ...mping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter c...

Страница 177: ...r Oscillation amplitude 1 Oscillation amplitude 1 Signal asymmetry Carrier mass flow Carrier pipe temperature Concentration Oscillation damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor...

Страница 178: ...ing 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation fre...

Страница 179: ...damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Excite...

Страница 180: ...sity Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1...

Страница 181: ...detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscil...

Страница 182: ...llation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Osc...

Страница 183: ...llation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Osc...

Страница 184: ...tion damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscill...

Страница 185: ...1 Measured values 1 Oscillation damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI P...

Страница 186: ...g 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequ...

Страница 187: ...ion damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscilla...

Страница 188: ...ping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation fr...

Страница 189: ...nsity Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1...

Страница 190: ...Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency...

Страница 191: ...ping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation fr...

Страница 192: ...mping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation f...

Страница 193: ...Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequenc...

Страница 194: ...damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation...

Страница 195: ...ing 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation fre...

Страница 196: ...1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current...

Страница 197: ...amping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter...

Страница 198: ...n damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Excit...

Страница 199: ...lation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Osci...

Страница 200: ...ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected...

Страница 201: ...llation damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1...

Страница 202: ...1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1...

Страница 203: ...Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1...

Страница 204: ...Oscillation frequency 1 Reference density Corrected volume flow Oscillation damping fluctuation 1 Oscillation damping fluctuation 1 Frequency fluctuation 1 Frequency fluctuation 1 Target mass flow Tem...

Страница 205: ...mping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation f...

Страница 206: ...damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation...

Страница 207: ...ng 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation freq...

Страница 208: ...flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation damping fluctuation 1 Oscillation damping fl...

Страница 209: ...mping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation f...

Страница 210: ...tion frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation damping fluctuation 1 Oscillation damping fluctuation 1 Frequency fluctuation 1 Frequency fluctuation 1 Tar...

Страница 211: ...Pulse output 1 to n 1 Check process 2 Check pulse output settings Measured variable status from the factory 1 Quality Good Quality substatus Ok Coding hex 0x80 to 0x83 Status signal S Diagnostic beha...

Страница 212: ...Reference density Corrected volume flow Oscillation damping fluctuation 1 Oscillation damping fluctuation 1 Frequency fluctuation 1 Frequency fluctuation 1 Target mass flow Temp compensated dynamic v...

Страница 213: ...Carrier mass flow Carrier pipe temperature Concentration Oscillation damping 1 Oscillation damping 1 Density Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscos...

Страница 214: ...equency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation damping fluctuation 1 Oscillation damping fluctuation 1 Frequency fluctuation 1 Frequency fluctuation 1 Target mas...

Страница 215: ...uency output Measured variable status Quality Good Quality substatus Ok Coding hex 0x80 to 0x83 Status signal C Diagnostic behavior Warning Diagnostic information Remedy instructions Influenced measur...

Страница 216: ...Good Quality substatus Ok Coding hex 0x80 to 0x83 Status signal C Diagnostic behavior Warning Diagnostic information Remedy instructions Influenced measured variables No Short text 497 Simulation blo...

Страница 217: ...nostic information Remedy instructions Influenced measured variables No Short text 594 Relay output simulation Deactivate simulation switch output Measured variable status Quality Good Quality substat...

Страница 218: ...ection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillat...

Страница 219: ...tection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscilla...

Страница 220: ...ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected...

Страница 221: ...ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected...

Страница 222: ...nematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation dampi...

Страница 223: ...inematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation damp...

Страница 224: ...ty Dynamic viscosity Sensor electronic temperature ISEM Empty pipe detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Os...

Страница 225: ...uency fluctuation 1 Target mass flow Temp compensated dynamic viscosity Temp compensated kinematic viscosity Temperature Status Volume flow Measured variable status Quality Good Quality substatus Ok C...

Страница 226: ...cillating 1 Check electronic 2 Inspect sensor Oscillation amplitude 1 Oscillation amplitude 1 Signal asymmetry Carrier mass flow Carrier pipe temperature Concentration Oscillation damping 1 Oscillatio...

Страница 227: ...tion Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillatio...

Страница 228: ...detection Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Osci...

Страница 229: ...ion Kinematic viscosity Low flow cut off Mass flow HBSI Pressure Exciter current 1 Exciter current 1 Oscillation frequency 1 Oscillation frequency 1 Reference density Corrected volume flow Oscillation...

Страница 230: ...g fluctuation 1 Oscillation damping fluctuation 1 Frequency fluctuation 1 Frequency fluctuation 1 Target mass flow Temp compensated dynamic viscosity Temp compensated kinematic viscosity Temperature S...

Страница 231: ...sage Previous diagnostics Two diagnostic events have already occurred Shows the diagnostic event that occurred prior to the current diagnostic event along with its diagnostic information Symbol for di...

Страница 232: ...ist F311 Electr failure Eventlist 0d01h19m10s A0014008 EN 36 Taking the example of the local display Max 20 event messages can be displayed in chronological order If the Extended HistoROM application...

Страница 233: ...I1079 Sensor changed I1089 Power on I1090 Configuration reset I1091 Configuration changed I1092 Embedded HistoROM deleted I1111 Density adjust failure I1137 Electronic changed I1151 History reset I115...

Страница 234: ...25 Write protection activated I1626 Write protection deactivated I1627 Web server login successful I1628 Display login successful I1629 CDI login successful I1631 Web server access changed I1632 Displ...

Страница 235: ...tile memory RAM to the factory setting e g measured value data The device configuration remains unchanged Restore S DAT backup Restore the data that are saved on the S DAT The data record is restored...

Страница 236: ...e Order code field Character string composed of letters numbers and certain punctuation marks e g Extended order code 1 Shows the 1st part of the extended order code The extended order code can also b...

Страница 237: ...the current version or the previous version using the service interface For the compatibility of the firmware version with the previous version the installed device description files and operating too...

Страница 238: ...13 2 Measuring and test equipment Endress Hauser offers a wide variety of measuring and test equipment such as W M or device tests Your Endress Hauser Sales Center can provide detailed information on...

Страница 239: ...M life cycle management database 14 2 Spare parts W M Device Viewer www endress com deviceviewer All the spare parts for the measuring device along with the order code are listed here and can be orde...

Страница 240: ...ing device and Connecting the measuring device sections in reverse order Observe the safety instructions 14 5 2 Disposing of the measuring device LWARNING Danger to personnel and environment from flui...

Страница 241: ...ed for the new transmitter WLAN antenna Wide range External WLAN antenna for a range of up to 50 m 165 ft Further information on the WLAN interface 82 Post mounting kit Post mounting kit for transmitt...

Страница 242: ...tform with online and on site tools Instant access for your staff to current in depth data shortens your plant s engineering time speeds up procurement processes and increases plant uptime Combined wi...

Страница 243: ...uids It can be used to read in the operating pressure value For details see Technical Information TI00383P and Operating Instructions BA00271P iTEMP The temperature transmitters can be used in all app...

Страница 244: ...ice remains in proper operating condition for its service life use the measuring device only for media against which the process wetted materials are sufficiently resistant 16 2 Function and system de...

Страница 245: ...150 4 6 0 to 550000 0 to 20240 Measuring ranges for gases The full scale values depend on the density of the gas and can be calculated with the formula below max G max F G x max G Maximum full scale...

Страница 246: ...red values are written from the automation system to the measuring device via the current input 246 Digital communication The measured values are written from the automation system to the measuring de...

Страница 247: ...passive Load 0 to 700 Resolution 0 38 A Damping Adjustable 0 07 to 999 s Assignable measured variables Mass flow Volume flow Corrected volume flow Density Reference density Temperature The range of o...

Страница 248: ...ses if the measuring device has one or more application packages Switch output Maximum input values DC 30 V 250 mA passive Open circuit voltage DC 28 8 V active Switching behavior Binary conductive or...

Страница 249: ...nment Choice of current output 4 to 20 mA active 0 4 to 20 mA passive Pulse frequency switch output Choice of current input 4 to 20 mA active 0 4 to 20 mA passive Status input The technical values cor...

Страница 250: ...and remedial measures Backlight Red backlighting indicates a device error Status signal as per NAMUR recommendation NE 107 Interface protocol Via digital communication PROFIBUS PA Via service interfa...

Страница 251: ...e flow Carrier mass flow Target mass flow Density Reference density Temperature Carrier pipe temperature Electronic temperature Current input The range of options increases if the measuring device has...

Страница 252: ...iting parameters is up to ten times faster with PROFIBUS upload download Condensed status Simplest and self explanatory diagnostic information by categorizing diagnostic messages that occur Configurat...

Страница 253: ...cross section 0 2 to 2 5 mm2 24 to 12 AWG Cable entries Cable gland M20 1 5 with cable 6 to 12 mm 0 24 to 0 47 in Thread for cable entry NPT G M20 Device plug for digital communication M12 Device plug...

Страница 254: ...ity specification only applies as of a minimum flow velocity of 0 2 m s in relation to the nominal diameter For the highest density measurement accuracy avoid significant tensile stresses due to the i...

Страница 255: ...Accuracy Max 50 ppm o r across the entire ambient temperature range Repeatability o r of reading 1 g cm3 1 kg l T medium temperature Base repeatability Mass flow and volume flow liquids 0 025 o r Mass...

Страница 256: ...ifference between the density calibration temperature and the process temperature the typical measured error of the sensor is 0 015 kg m3 C 0 0075 kg m3 F kg m 3 0 0 5 1 0 1 5 2 0 2 5 3 0 C F 200 150...

Страница 257: ...tion of the flow rate Flow rate Maximum measured error in o r ZeroPoint BaseAccu 100 A0021332 BaseAccu A0021339 ZeroPoint BaseAccu 100 A0021333 ZeroPoint MeasValue 100 A0021334 Calculation of the maxi...

Страница 258: ...play module IP20 type 1 enclosure Sensor As standard IP66 67 type 4X enclosure External WLAN antenna IP67 Vibration resistance Vibration sinusoidal according to IEC 60068 2 6 2 to 8 4 Hz 3 5 mm peak 8...

Страница 259: ...ment Secondary containment The sensor housing is filled with helium and protects the electronics and mechanics inside The following secondary containment pressure ratings burst pressures are only vali...

Страница 260: ...r judges that the secondary containment pressure rating burst pressure does not provide an adequate safety margin the device can be fitted with a rupture disk This will prevent extensive pressure buil...

Страница 261: ...st stainless 15 6 kg 34 4 lbs Sensor Sensor with aluminum connection housing version see the information in the following table Cast connection housing version stainless 3 7 kg 8 2 lbs Weight in SI un...

Страница 262: ...Hygienic version for maximum corrosion resistance stainless steel 1 4404 316L Option L Cast stainless 1 4409 CF3M similar to 316L Cable entries cable glands 1 2 3 4 A0028352 37 Possible cable entries...

Страница 263: ...old plated brass Connecting cable Connecting cable for sensor Proline 500 digital transmitter PVC cable with copper shield Connecting cable for sensor Proline 500 transmitter Standard cable PVC cable...

Страница 264: ...ian Dutch Portuguese Polish Russian Turkish Chinese Japanese Korean Arabic Bahasa Indonesian Thai Vietnamese Czech Swedish Via Web browser English German French Spanish Italian Dutch Portuguese Polish...

Страница 265: ...ce 271 DeviceCare SFE100 Notebook PC or tablet with Microsoft Windows system CDI RJ45 service interface WLAN interface Fieldbus protocol 242 FieldCare SFE500 Notebook PC or tablet with Microsoft Windo...

Страница 266: ...evice memory This memory can be overwritten with an updated data record for example after commissioning Additional information on the data storage concept There are different types of data storage uni...

Страница 267: ...s e g DeviceCare FieldCare or Web server Data logging Manual If the Extended HistoROM application package order option is enabled Record up to 1000 measured values via 1 to 4 channels User configurabl...

Страница 268: ...e RED 2014 53 EU United States of America CFR Title 47 FCC Part 15 247 Canada RSS 247 Issue 1 Japan Article 2 clause 1 item 19 Additional country specific approvals on request Additional certification...

Страница 269: ...nt of a power failure in field and control instruments with microprocessors NAMUR NE 43 Standardization of the signal level for the breakdown information of digital transmitters with analog output sig...

Страница 270: ...er 7 6 a Control of monitoring and measuring equipment Functional testing in the installed state without interrupting the process Traceable verification results on request including a report Simple te...

Страница 271: ...device parameters Measuring device Documentation code Promass 500 GP01061D Supplementary device dependent documentation Safety Instructions Contents Documentation code Measuring device ATEX IECEx Ex...

Страница 272: ...l data Proline Promass Q 500 PROFIBUS PA 272 Endress Hauser Installation Instructions Contents Documentation code Installation Instructions for spare part sets Overview of accessories available for or...

Страница 273: ...device Proline 500 46 Proline 500 digital 39 Connecting the signal cable supply voltage cable Proline 500 digital transmitter 44 Proline 500 transmitter 50 Connection see Electrical connection Connect...

Страница 274: ...le 86 Establishing a connection 84 Function 83 User interface 84 Filtering the event logbook 233 Firmware Release date 86 Version 86 Firmware history 237 Flow direction 23 28 Flow limit 260 Function c...

Страница 275: ...n view 64 Navigation view In the submenu 64 In the wizard 64 Numeric editor 66 O Operable flow range 246 Operating elements 67 167 Operating keys see Operating elements Operating menu Menus submenus 6...

Страница 276: ...rating conditions 253 Registered trademarks 9 Remedial measures Calling up 168 Closing 168 Remote operation 265 Repair of a device 239 Repairs 239 Notes 239 Repeatability 255 Replacement Device compon...

Страница 277: ...asured variable 63 For measurement channel number 63 For menus 65 For parameters 65 For status signal 63 For submenu 65 For wizard 65 In the status area of the local display 63 In the text and numeric...

Страница 278: ...ine access code 141 Display 125 Low flow cut off 128 Partially filled pipe detection 129 Pulse frequency switch output 113 115 120 Relay output 1 to n 122 Select medium 104 WLAN settings 138 Workplace...

Страница 279: ......

Страница 280: ...www addresses endress com...