3.5.2.2 Setting up modbus communications - additional option
NOTE:
Modbus is optional and this section is applicable only
if the modbus option was ordered.
The
Sentinel LCT8
supports digital communications using
the
MODBUS/RTU
protocol, with 2-wire RS-485 or 3-wire
RS-232C as the physical layer interfaces. Baud rate can be
specified from 4800 to 19200 bits per second (bps), with
selectable parity and number of stop bits (default =
9600
bps,
even
parity and
1
stop bit).
To set up modbus communication settings, refer to
Figure
55
below and complete the steps on the next page.
NOTE:
If necessary, refer to
as a
reminder about color code meanings.
Figure 55: Communication - modbus setup
Legend
Green
At least user 3 level access required to follow the step.
Red
At least user 1 access required to follow the step.
Red
Option only visible to user 1 or above.
Green
Option only visible to user 3 or above.
Use
or
arrow
to highlight the
preferred”option
and press
twice to confirm
selection.
User1 and user2
access have
only “stop bits”
option visible.
So, the
or
arrow steps are
not applicable to
these users.
Use
arrow to highlight
“modbus” if not already
highlighted.
46
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...