1. Refer to
and navigate to
Page 3: program
.
2. Use the [
] and [
] keys to highlight the desired channel.
Press the [
] key twice to confirm your choice.
3. Use the [
] and [
] keys to highlight the
error limits
option. Then, press the [
] key to confirm your selection
or press the [
] key to return to the previous screen.
4. Use the [
] and [
] keys to highlight the desired
signal parameter
. Then, press the [
] key to confirm
your selection or press the [
] key to return to the
previous screen.
5. Enter the desired
error limit
for the chosen signal
parameter. Press the [
] key to confirm the new value
or press the [
] key to keep the original value. In either
case, you will then return to the previous screen.
6. Repeat
steps 4 and 5
as necessary, to program
additional signal parameter error limits.
3.4.1.4 Understanding the error limits
Note the following information about the
Sentinel LCT8
error limits:
• The
signal min
limit is the minimum acceptable value
for the transducer signal received by the
Sentinel LCT8
.
If the signal falls below the programmed limit, the
E1: LOW SIGNAL
error is displayed.
• The measured velocity value is expected to be
within the
min velocity
and
max velocity
limits. If the
measured velocity is outside the programmed range,
a
W3: VELOCITY RANGE
warning is displayed.
• The amplitude of the received signal is expected to
be within the
amplitude min
and
amplitude max .
programmed limits. If the signal falls outside these limits,
the
E5: AMPLITUDE
error is displayed.
• If the fluid soundspeed exceeds that entered in the
normal mode of fluid
submenu by more than the
percentage programmed in the
NormalMde SoS
error limit, the
E2: SOUNDSPEED
error is displayed.
• If the velocity changes by more than the programmed
acceleration
error limit from one measurement to the
next, the
E6: ACCELERATION
error is displayed.
• If the SNR is below the programmed
SNR
error limit, the
W24:SNR
warning is displayed.
NOTE:
Refer to
for more details about errors and
warnings.
32
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...