5.7.3 Alarm output test
To test an
alarm output
, refer to
Figure 96
below and complete the following steps.
NOTE:
If necessary, refer to
as a reminder about color code meanings.
1. Refer to
and navigate to
page 1: manufacturing tests .
2. Use the [
] and [
] keys to highlight
IO test
and press the
[
] key to confirm your choice.
3. Use the [
] and [
] keys to highlight
alarm test
and press
the [
] key to confirm your choice or press the [
] key to
return to the previous screen.
4. To test the desired
alarm output
, use the [
] and [
] keys
to highlight either alarm #1 or alarm #2. Then, press the
[
] key to confirm your choice or press the [
] key to
return to the previous screen.
5. With
alarm state
highlighted, press the [
] key to
confirm the selection or press the [
] key to return to the
previous screen.
6. To select the alarm state you wish to test, use the [
] and
[
] keys to highlight either
trip or reset
and press the [
]
key twice to confirm your choice or press the [
] key to
cancel and return to the previous screen.
Legend
Options in
requires at least user admin access
Options in
requires at least user 1 access
Options in
requires at least user 3 access
NOTE:
To perform IO test on
alarm #2 highlight the option
before
key is pressed.
Press
Press
or
Use
or
arrow to highlight the
“alarm state” option you want to
test and press
twice to confirm
selection.
Figure 96: Manufacturing IO test - alarm output
102
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...