3.5 Generic meter configuration
This section describes the procedures for configuring
the following:
•
Units
(see
below
)
•
Communication
)
•
Reset totals
Setting reset totals on page 50
•
Totalize on error
Setting up totalizer error handling
)
•
Date/time
Setting the date and time on page 52
)
3.5.1 Selecting the preferred unit type
To select the desired units, refer to
Figure 53
below and
complete the following steps:
NOTE:
If necessary, refer to
as a
reminder about color code meanings.
1. Refer to
and navigate to
page 4: configuration (generic) .
2. Use the [
] and [
] keys to highlight
units
and press the
[
] key to confirm your choice.
3. At the next screen, use the [
] and [
] keys to highlight
either the
metric or English
option. Then, press the [
]
key twice to confirm your selection or press the [
] key
to cancel and return to the previous screen.
Figure 53: Unit type selection
Use
or
arrow to highlight the
preferred “units” option and press
twice to confirm selection.
Legend
Green
At least user 3 level access required to follow the step.
Dotted
green
Not all transitional screens are shown (at least user 3 level
access required).
Red
Option only visible to user 1 or above.
Green
Option only visible to user 3 or above.
43
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...