5.8.1.2 Formatting and viewing audit log records
The
audit log
records have neatly formatted columns for easy readability. If the record types listed in
contain irrelevant/not applicable column values, they are represented as “---” in the corresponding record column. A typical
audit log file record appears as follows:
Rec#
Date
Time
Source
ID
Name
Auth
Old
New
1
12/11/2008
2:13:19
Meter
---
Power
---
---
Power-on
2
12/11/2008
2:13:22
---
1
Alarm
---
---
Tripped
Similarly, a parameter change record would appear
as follows:
Rec#
Date
Time
Source
ID
Name
Auth
Old
New
1
12/11/2008
5:11:58
1
25798
RxW width
0
1024
2048
The user may view the meter’s
audit log file
on the PC screen or may save it as either a text file or a
Microsoft Excel
file. To get
the full benefits of the
Sentinel LCT8
audit trail
feature, we recommend using the Microsoft Excel file option, which is available
only if Microsoft® Excel 2000 or later is installed on the PC. Additional formatting may be applied in the Microsoft Excel file for
enhanced readability.
5.8.2 Error logs
The
Sentinel LCT8
error log
records system error events in
a user-readable log file and stores that file in persistent
memory. The error event is also recorded in the
audit log
file, whenever the meter changes from a No error state to an
error state or vice versa.
The meter can store up to 250 error event records in the
error log
file. When the file size reaches this limit, any new
record will overwrite the oldest existing record in the file. This
is called a
circular log
, and it ensures that the user is always
viewing the 1000 most recent event records.
The
error log
file is not viewable on the
Sentinel LCT8
display, and the contents of the file stored in the meter’s
memory cannot be altered. However, the
error log
file can
be downloaded to a PC using the Panametrics PanaView
SEN898 software. The log file saved to the PC opens as a text
file, and a typical
error log
file record appears as follows:
Rec#
Date
Time
Source
Name
OldBitmap
Old
NewBitmap
New
1
12/12/2016
14:36:35
9
Error
0x0
E0: No error
0xf000001
E25: Half full pipe
104
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...