3.4.1.2 Entering the signal parameters
Use the signal submenu, to set the
Sentinel LCT8
transducer
signal parameters.
IMPORTANT
The factory default
signal parameters
are suitable for
most applications. Consult the factory before changing
any of these settings.
To configure the
signal parameters
, refer to
Figure 44
below
and complete the following steps:
NOTE:
If necessary, refer to
as a
reminder about color code meanings.
1. Refer to
and navigate to
Page 3: program .
2. Use the [
] and [
] keys to highlight the desired channel.
Press the [
] key twice to confirm your choice.
3. Use the [
] and [
] keys to highlight the signal option.
Then, press the [
] key to confirm your selection or press
the [
] key to return to the previous screen.
4. Use the [
] and [
] keys to highlight the zero cutoff or
the active tw peak pcnt option. Then, press the [
] key to
confirm your selection or press the [
] key to return to
the previous screen.
NOTE:
The zero cutoff parameter causes fluid velocities less
than the setpoint to be reported as zero.
5. Enter the desired value for the signal parameter chosen
in the previous step. Press the [
] key to confirm the
new value or press the [
] key to keep the original value.
You will then return to the previous screen.
Figure 44: Entering the signal parameters
Legend
Red
At least user 1 level access required to follow the step.
Green
At least user 3 level access required to follow the step.
Green
Option only visible to user 3 or above.
NOTE:
To select channel
2 to channel 8, highlight
the option and press
key twice to confirm
the selection. Once you
confirm your selection
you will see the next
screen.
The
arrow on the
screen usually indicates
there are more options
below and the
arrow
can be used to scroll to
those options.
Use
or
arrow to
highlight the “signal…”
if not already highlighted.
or
or
30
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...