5.3 Flow info
The
Sentinel LCT8
offers built-in diagnostic parameters to aid in troubleshooting. To view multiple diagnostics for all channels
via the flow info form, refer to
Figure 92
below and complete the steps on the next page.
NOTE:
If necessary, refer to
as a reminder about color code meanings.
Figure 92: Using flow info in troubleshooting
Use
or
arrow to view these (p1 to p7) diagnostics for other
channels (ch1 to ch8).
This page provides “active tw” diagnostics. Use
or
arrow to
view “active tw” diagnostics for other channels (ch1 to ch8).
This page provides “current errors” diagnostics. This page displays
all errors on a specific channel at any point in time. Use
or
arrow to view “current errors” for other channels (ch1 to ch8).
This page provides “error bitmap” diagnostics. This page displays
all errors on a specific channel at any point in time and past 15
measurement cycles. Use
arrow to view page2 or the “error
bitmap” diagnostics. Use
or
arrow to view these diagnostics for
other channels (ch1 to ch8).
This page provides “tracking” diagnostics. This page displays
diagnostics related to tracking the signal. This page is helpful
only when the meter is in error. Use
or
arrow to view “tracking”
diagnostics for other channels (ch1 to ch8).
NOTE:
The arrows in this figure are red and indicate at least user1 access level is required to follow these steps.
NOTE:
The flow info pages are updated
with live measurement diagnostics and
hence constantly varying.
94
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...