3.6.3.1 Understanding the alarm type selection
If
normal
or
failsafe
was selected as the
alarm type
, you must then program the parameters listed in
Table 11
below.
Table 11: Alarm type options
Options
Parameters
Description
Level
Units
This defines the measured units monitored by the alarm
Trip point
This defines the measured value required to trip the alarm
Dead band
This defines a range of values around the trip point which neither trip nor
reset the alarm.
Trip direction
The alarm may be set to trip either when the value drops below the trip point
(low alarm) or when the value moves above the trip point (high alarm).
Error level
Choose from the following options:
Reset
- Forces the alarm to reset on an error or fault condition.
Trip
- Forces the alarm to trip on an error or fault condition.
Hold last
- Holds the current alarm status on an error or fault condition.
Direction
Units
The alarm can be tripped on either forward or reverse flow, using volumetric
flow in m
3
/s of the selected channel. The alarm determines direction by
looking for a positive or negative change in the flow reading.
Fault
Alarm units
The alarm is set to trip based on the
channel
selected here.
Flow
This trips the alarm when a flow error occurs on the selected channel.
Non-flow
This trips the alarm when a non-flow error occurs on the selected channel.
All
This trips the alarm on any error on the selected channel.
Custom
The user can select the errors that should trip the alarm.
3.6.3.2 Configuring the alarm latch state
To configure the alarm latch state options, see
Table 12
below.
Table 12: Latch state options
Option
Description
Latch
This enables the alarm to hold its current state, by causing the
user intervene indicator
to be lit when
the alarm is tripped.
No Latch
The alarm is reset when the fault or error condition clears.
Clear latch
This changes the latched state, by turning off the
user intervene indicator
when the fault or error
condition clears.
62
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...