1. Refer to
and navigate to
page 5: IO
.
2. Use the [
] and [
] keys to highlight alarms and press
the [
] key to confirm your choice.
3. Use the [
] and [
] keys to highlight either alarm #1
or alarm #2. Then, press the[
] key to confirm your
selection or press the [
] key to cancel and return to the
previous screen.
NOTE:
As the two alarms have similar options, only the set up
of alarm #1 is described here, as an example.
4. Use the [
] and [
] keys to highlight alarm status and
press the [
] key to confirm your selection. Then, use
the [
] and [
] keys to highlight the desired option (off,
normal or failsafe) and press the [
] key twice to confirm
your selection or press the [
] key to cancel and return
to the previous screen.
NOTE:
Off = alarm is disabled, normal = alarm is normally
de-energized and is energized when tripped, failsafe =
alarm is normally energized and is de-energized when
tripped. See “Understanding the alarm type Selection” on
page 77 for more information.
5. Use the [
] and [
] keys to highlight type and press the
[
] key to confirm your selection. Then, highlight and
configure each of the following options (see Table 11 on
page 77):
a.
Level
- program the units, trip point, dead band, trip
direction, and error level. Then, press the[
] key to
confirm your settings or press the [
] key to cancel
and return to the previous screen.
b.
Direction
- program the units. Then, press the [
]
key to confirm your selection or press the [
] key to
cancel and return to the previous screen.
c.
Fault
- Select from alarm units, flow, Non-flow, All,
and custom. Then, press the [
] key to confirm your
settings or press the [
] key to cancel and return to
the previous screen.
6. Use the [
] and [
] keys to highlight Latch state and
press the [
] key to confirm your selection. Then, use the
[
] and [
] keys to highlight the desired option (Latch,
No Latch or Clear Latch) and press the [
] key twice to
confirm your selection or press the [
] key to cancel and
return to the previous screen.
NOTE:
See “configuring the alarm Latch state” on page 77 for
a description of the latch state options.
7. When you have finished, press the [
] key to confirm
your selections or press the [
] key to cancel and return
to the previous screen.
61
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...