5.7 Manufacturing tests
The procedures for testing certain key functionalities of the
Sentinel LCT8
are described in this section. The
Sentinel
LCT8
Watchdog test, keypad test, display test and IO test are
available to the user. As an example, the IO test is discussed
in detail in this chapter. All other manufacturing tests are
performed in a similar manner.
The IO test is used to troubleshoot any problems associated
with damaged electronics or changes in the process
conditions that may be affecting the
Sentinel LCT8
inputs
and/or outputs. Proceed to the following three sections for
instructions on how to test specific IO functions.
5.7.1 Frequency output test
To test a frequency output, refer to
Figure 94
below and
complete the steps on the following page.
NOTE:
If necessary, refer to
as a
reminder about color code meanings.
Figure 94: Manufacturing IO test - frequency output
Use
or
arrow to highlight
the “drive control” option you
want and press
twice to
confirm selection.
Legend
Options in
requires at least user admin access
Options in
requires at least user 1 access
Options in
requires at least user 3 access
NOTE:
To perform io test on
freq/tot #2 choose highlight the
option before
key is pressed.
Use
or
arrow keys to
position the cursor.
Use
or
arrow keys to
increment or decrement the
digits to the “test frequency”
you would like to see on the
frequency output.
98
Содержание Panametrics Sentinel LCT8
Страница 1: ...Sentinel LCT8 User s manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 29: ...Figure 32 Wiring LCT8 to SEN898 electronics cross plane 6 24 flowcell local mount ref dwg 702 2041 19...
Страница 30: ...20...
Страница 98: ...88...
Страница 122: ...no content intended for this page 112...
Страница 132: ...122...