5.7.2 Totalizer output test
To test a
totalizer output
, refer to
Figure 95
below and complete the steps on the next page.
NOTE:
If necessary, refer to
as a reminder about color code meanings.
Figure 95: Manufacturing IO test - totalizer output
Legend
Options in
requires at least user admin access
Options in
requires at least user 1 access
Options in
requires at least user 3 access
Use
or
arrow to highlight the
“polarity” option you want to test
and press
twice to confirm
selection.
Use
or
arrow keys to position
the cursor.
Use
or
arrow keys to
increment or decrement the
digits to the “pulse count” for
totalizer test.
Use
or
arrow to highlight the
option to test.
Note: To perform io test on
freq/tot #2 choose highlight the
option before
key is pressed.
Use
or
arrow keys to position
the cursor.
Use
or
arrow keys to
increment or decrement the
digits to the “pulse width” for
totalizer test.
Use
or
arrow to highlight the
“drive control” option you want to
test and press
twice to confirm.
100