1

Bluetooth

®

LED will blink red.

5. Launch Graphical Analysis 4.
6. Click or tap Sensor Data

Collection.

7. Click or tap your Go Direct sensor

from the list of Discovered
Wireless Devices. Your sensor's ID
is located near the barcode on the
sensor. The Bluetooth LED will
blink green when it is successfully
connected.

8. Click or tap Done to enter data-

collection mode.

9. For best results, perform a two-

point calibration using the High
and Low Standard solutions.

4. Launch Graphical Analysis 4 or turn

on LabQuest 2.

5. For best results, perform a two-point

calibration using the High and Low
Standard solutions.

Charging the Sensor

Connect the Go Direct Ammonium Ion-Selective Electrode to the included
Micro USB Cable and any USB device for two hours. Connecting the Go Direct
Ammonium Electrode BNC to the amplifier during charging is optional.

You can also charge up to eight Go Direct Ammonium Ion-Selective Electrodes
using our Go Direct Charge Station, sold separately (order code: GDX-CRG).
An LED on each Go Direct Ammonium Ion-Selective Electrode indicates
charging status.

Charging

Blue LED on steady while sensor is connected to
the Micro USB Cable or Charge Station.

Fully charged

Blue LED is off when charging is complete.

Go Direct

®

Ammonium Ion-
Selective Electrode

(Order Code GDX-NH4)

Go Direct Ammonium Ion-Selective Electrode (ISE) is used to measure the
concentration of ammonium (NH

4

+

) ions in aqueous samples.

Note:

Vernier products are designed for educational use. Our products are not

designed nor are they recommended for any industrial, medical, or commercial
process such as life support, patient diagnosis, control of a manufacturing
process, or industrial testing of any kind.

What's Included

l

Go Direct Ammonium Ion-Selective Electrode (Go Direct Ion-Selective
Electrode Amplifier connected to a Go Direct Ammonium Ion-Selective
Electrode BNC Electrode)

l

30 mL bottle of High Standard solution with SDS (100 mg/L NH

4

+

-N)

l

30 mL bottle of Low Standard solution with SDS (1 mg/L NH

4

+

-N)

l

Short-Term ISE Soaking Bottle

l

Micro USB Cable

Compatible Software

See

www.vernier.com/manuals/gdx-nh4

for a list of software compatible with the

Go Direct Ammonium Ion-Selective Electrode.

Getting Started

Please see the following link for platform-specific connection information:

www.vernier.com/start/gdx-nh4

Bluetooth Connection

USB Connection

1. Install Graphical Analysis 4 on

your computer, Chromebook™, or
mobile device. See
www.vernier.com/ga4 for software
availability.

2. Charge your sensor for at least

2 hours before first use.

3. Prepare the electrode by soaking it

in the High Standard solution for
30 minutes. Refer to the Using the
Product section for more
information.

4. Turn on your sensor by pressing

the power button once. The

1. Prepare the electrode by soaking it

in the High Standard solution for
30 minutes. Refer to the Using the
Product section for more
information.

2. Install Graphical Analysis 4 on your

computer or Chromebook. If using
LabQuest 2, make sure LabQuest
App is up to date. See
www.vernier.com/ga4 for software
availability or
www.vernier.com/downloads to
update LabQuest App.

3. Connect the sensor to the USB port.

Содержание Go Direct GDX-NH4

Страница 1: ...Our products are not designed nor are they recommended for any industrial medical or commercial process such as life support patient diagnosis control of a manufacturing process or industrial testing of any kind What s Included l Go Direct Ammonium Ion Selective Electrode Go Direct Ion Selective Electrode Amplifier connected to a Go Direct Ammonium Ion Selective Electrode BNC Electrode l 30 mL bot...

Страница 2: ... on the sensor Press button once Red LED indicator flashes when unit is on Putting the sensor in sleep mode Press and hold button for more than three seconds to put into sleep mode Red LED indicator stops flashing when sleeping Connecting the Sensor See the following link for up to date connection information www vernier com start gdx nh4 Connected and charging Blue and Green LED solid when sensor...

Страница 3: ...lute the High Standard by a factor of 100 Replacement Modules The Go Direct Ammonium Ion Selective Electrode has a PVC membrane module with a limited life expectancy The module is warranted to be free from defects for a period of 1 year from the date of purchase It is possible however that a membrane module will work well after the warranty period If you notice a reduced response then it is probab...

Страница 4: ...nds on the battery Although the battery is warranted for one year the expected battery life should be several years Replacement batteries are available from Vernier order code GDX BAT 300 Water Resistance The Go Direct Ammonium Ion Selective Electrode is not water resistant and should never be immersed in water above the BNC junction If water gets into the device immediately power the unit down pr...

Страница 5: ...ndicates that this product must not be disposed of in a standard waste container Troubleshooting Using Ionic Strength Adjustor ISA Solutions to Improve Accuracy For optimal results at low concentrations of ions a standard method for making measurements with ion selective electrodes is to add ionic strength adjustor ISA solutions to each of your standard solutions and samples Adding an ISA ensures ...

Страница 6: ...mpt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Industry Canada Class B This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference causin...

Отзывы: