Vernier Go Direct GDX-NH4 Скачать руководство пользователя страница 4

4

Ion Examples

n (valence of

ion)

S (at 25 °C),

mV/decade

Calcium (Ca

2+

)

+2

+29.58

Potassium (K

+

), Ammonium (NH

4

+

)

+1

+59.16

Nitrate (NO

3

-

), Chloride (Cl

-

)

–1

–59.16

Assuming

C

0

is near zero, the equation can be rewritten as:

C

= 10

˄

[(

E – E

o

) /

S

]

allowing for the calculation of the ion concentration.

It is very important to note that this table reflects ideal behavior. Ion-selective
electrodes have slopes that are typically lower than ideal. It is generally
accepted that an ISE slope from 88–101% of ideal is allowable. The slope (S) is
an indicator of ISE performance. If the slope changes significantly over time, it
may indicate that it is necessary to replace the ISE sensor tip.

Potential

vs.

Concentration

To measure the mV readings from an aqueous sample, calibration is not
required. To convert mV readings to concentration (mg/L or ppm), the software
uses a modified version of the Nernst Equation:

C

= 10

˄

[(

E – E

o

) /

S

m

]

C

= concentration of ion to be measured (mg/L or ppm)

E

= measured potential of sample (mV)

E

o

= measured potential (mV) at a C = 1 mg/L Ca

2+

concentration

S

m

= measured electrode slope in mV/decade

The value of

S

m

, the measured electrode slope, is determined by measuring the

potential of two standard solutions, and solving the equation below:

S

m

= – [(Low Standard – High Standard) / # of decades*]

Example Calculation, converting mV to mg/L

For this example, the measured quantities are shown in the chart below:

Solution

Measured Potential

1 mg/L NH

4

+

standard

0 mV

100 mg/L NH

4

+

standard

116 mV

unknown sample

88 mV

C = 10^[(88 mV – 0 mV) / 58 mV/decade] = 33 mg/L NH

4

+

-N

Battery Information

The Go Direct Ammonium Ion-Selective Electrode contains a small lithium-ion
battery in the handle. The system is designed to consume very little power and
not put heavy demands on the battery. Although the battery is warranted for one
year, the expected battery life should be several years. Replacement batteries are
available from Vernier (order code: GDX-BAT-300).

Water Resistance

The Go Direct Ammonium Ion-Selective Electrode is not water resistant and
should never be immersed in water above the BNC junction.

If water gets into the device, immediately power the unit down (press and hold
the power button for more than three seconds). Disconnect the sensor and
charging cable, and remove the battery. Allow the device to dry thoroughly
before attempting to use the device again. Do not attempt to dry using an
external heat source.

How the Sensor Works

Combination Ion-Selective Electrodes consist of an ion-specific (sensing) half-
cell and a reference half-cell. The ion-specific half-cell produces a potential that
is measured against the reference half-cell depending on the activity of the
target ion in the measured sample. The ion activity and the potential reading
change as the target ion concentration of the sample changes. The relationship
between the potential measured with the ISE and the ion activity, and thereby
the ion concentration in the sample, is described by the Nernst equation:

E

= measured potential (mV) between the ion-selective and the reference

electrode

E

o

= standard potential (mV) between the ion-selective and reference electrodes

R = universal gas constant (R = 8.314 J mol

-1

K

-1

)

T

= temperature in K (Kelvin), with T (K) = 273.15 + t °C where

t

is the

temperature of the measured solution in °C.

F = Faraday constant (96485 C mol

-1

)

n

= valence of the ion

C

= concentration of ion to be measured

C

o

= detection limit

Since R and F are constant, they will not change. The electrical charge of the
ion (valence) to be measured is also known. Therefore, this equation can be
simplified as:

E = E

o

–S

log(C + C

o

)

where

is the ideal slope of the ISE.

The following table describes ideal behavior:

Содержание Go Direct GDX-NH4

Страница 1: ...Our products are not designed nor are they recommended for any industrial medical or commercial process such as life support patient diagnosis control of a manufacturing process or industrial testing of any kind What s Included l Go Direct Ammonium Ion Selective Electrode Go Direct Ion Selective Electrode Amplifier connected to a Go Direct Ammonium Ion Selective Electrode BNC Electrode l 30 mL bot...

Страница 2: ... on the sensor Press button once Red LED indicator flashes when unit is on Putting the sensor in sleep mode Press and hold button for more than three seconds to put into sleep mode Red LED indicator stops flashing when sleeping Connecting the Sensor See the following link for up to date connection information www vernier com start gdx nh4 Connected and charging Blue and Green LED solid when sensor...

Страница 3: ...lute the High Standard by a factor of 100 Replacement Modules The Go Direct Ammonium Ion Selective Electrode has a PVC membrane module with a limited life expectancy The module is warranted to be free from defects for a period of 1 year from the date of purchase It is possible however that a membrane module will work well after the warranty period If you notice a reduced response then it is probab...

Страница 4: ...nds on the battery Although the battery is warranted for one year the expected battery life should be several years Replacement batteries are available from Vernier order code GDX BAT 300 Water Resistance The Go Direct Ammonium Ion Selective Electrode is not water resistant and should never be immersed in water above the BNC junction If water gets into the device immediately power the unit down pr...

Страница 5: ...ndicates that this product must not be disposed of in a standard waste container Troubleshooting Using Ionic Strength Adjustor ISA Solutions to Improve Accuracy For optimal results at low concentrations of ions a standard method for making measurements with ion selective electrodes is to add ionic strength adjustor ISA solutions to each of your standard solutions and samples Adding an ISA ensures ...

Страница 6: ...mpt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Industry Canada Class B This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference causin...

Отзывы: