USING THE SLIDER LASER TIRF WITH CUSTOMERS' OWN LASERS
Carl Zeiss
Remote switch for laser mode
Laser TIRF 3
96
423683-7144-001
02/2014
6.3
Remote switch for laser mode
The Laser TIRF imaging system can be equipped optionally with the remote switch for laser mode which
has to be connected between the Slider Laser TIRF and the distribution box for laser safety using a Y-
cable.
Laser operation mode is activated on the remote control switch by means of the laser control button
(Fig. 6-4/
1
) (button depressed;
Active
3
)
lights up) or deactivated (button not depressed;
Active
LED not lit).
If laser operation mode is activated, the eyepiece shutter is closed.
The
Safe
2
) indicates the status of the interlock connections. The
Safe
LED lights up only if
all interlock connections of the laser TIRF imaging system are closed. In this case the laser shutter opens
automatically.
The Slider Laser TIRF is equipped with identical
Active
and
Safe
LEDs which also display the
status of the laser operation mode and the interlock connections.
Fig. 6-4
Remote switch for laser mode
For connecting the remote control switch please see Fig. 6-5 on page 97.
Independently of the remote switch for laser mode, the laser operation mode can also be
activated / deactivated on the button ring of the microscope or using the software.