USING THE SLIDER LASER TIRF WITH CUSTOMERS' OWN LASERS
Carl Zeiss
Safety Check Prior to First Start-up
Laser TIRF 3
104
423683-7144-001
02/2014
Action
Active LED Safe LED
System status
Transmitted light carrier of the stand
pivoted out
lights up
does not
light up
•
Laser shutter closed
•
Eyepiece shutter open
Transmitted light carrier of the stand
pivoted in
lights up
lights up
•
Laser shutter open
•
Eyepiece shutter closed
Distribution box for laser safety not
connected to microscope
lights up
does not
light up
•
Laser shutter closed
•
Eyepiece shutter open
Not all sockets of the distribution box
for laser safety occupied
lights up
does not
light up
•
Laser shutter closed
•
Eyepiece shutter open
Camera or protective cover removed
from the camera adapter
lights up
does not
light up
•
Laser shutter closed
•
Eyepiece shutter open
Reflected light lamp (e.g. HXP 120) or
the protective cover are not on the
reflected light illumination connector
of the microscope
lights up
does not
light up
•
Laser shutter closed
•
Eyepiece shutter open
The camera or protective cover is not
on the camera adapter.
lights up
lights up
•
Laser shutter open
•
Eyepiece shutter closed
Reflected light lamp (e.g. HXP 120) or
the protective cover are not on the
reflected light illumination connector
of the microscope
lights up
lights up
•
Laser shutter open
•
Eyepiece shutter closed
If the safety system does not work in the way described above, take the instrument out of use,
label it clearly as being out of order and immediately contact Carl Zeiss Microscopy GmbH
Service.
Attach the microscope tube again after successful safety check.