6.9
Power Requirements
Three wall outlets at a maximum distance of 1.5 m from the Laser Bench Tower or Rack are required to
connect the system.
Use the three multiple socket-outlets, supplied, to connect the system components (see Fig. 5.10).
The multiple socket-outlets must not be connected in series!
Fig. 6-10
Terminal assignments of the multipoint connectors
The Laser TIRF imaging system must be connected to the power supply using several multiple
sockets connected separately to installation sockets. Connection via a single multiple socket
plugged into a single installation socket would lead to inadmissibly high contact current levels.
The customer is responsible for ensuring that the permissible contact current levels are not
exceeded.