START-UP
Carl Zeiss
Installation Requirements
Laser TIRF 3
52
423683-7144-001
02/2014
3
START-UP
The Laser TIRF system is a class 3B laser system and labeled accordingly (total output power:
max. 500 mW). If you use laser instruments, adhere to the provisions concerning health
protection and industrial safety. Users should also ensure that they observe all relevant legal
requirements.
Inform the laser safety officer about the intended first start-up of the Laser TIRF imaging
system.
The Laser TIRF imaging system has to be installed or retrofitted and adjusted in all cases by the
Carl Zeiss Microscopy GmbH Service department or persons authorized by the manufacturer. If
used as intended (see Section 2.1 on page 23), the Laser TIRF imaging system can be started up
and used by the customer without additional precautions.
Before installing and starting up the instrument it is imperative to read
carefully.
3.1
Installation Requirements
3.1.1
Space Requirements
The following system installation requirements have to be taken into consideration:
−
Floor area required 7 m
2
,
−
stable laboratory worktable with system on vibration-isolated base plate or vibration-isolated, air-
cushioned microscope table (with dimensions of 180 x 90 cm),
−
Minimum space between system and wall 50 cm.