INSTRUMENT DESCRIPTION
Carl Zeiss
Technical Data
Laser TIRF 3
50
423683-7144-001
02/2014
SVB 1 (signal distribution box)
Line voltage ......................................................................................................... 100 V to 240 V
±
10 %
Change of line voltage is not required!
Line frequency ......................................................................................................................... 50/60 Hz,
Max. power consumption .............................................................................................................. 20 VA
Fuses .................................................................................................................................. 2x T 1.0 A/H
Environment for use ........................................................................................................... Closed rooms
Protection class ..................................................................................................................................... I
Ingress protection rating .................................................................................................................. IP 20
Electrical safety ........................................................... in compliance with DIN EN 61010-1 (IEC 61010-1)
including CSA and UL directives
Overvoltage category............................................................................................................................. II
TIRF controller with desktop power supply
Line voltage ......................................................................................................... 100 V to 240 V
±
10 %
Change of line voltage is not required!
Line frequency ......................................................................................................................... 50/60 Hz,
Power consumption .................................. ~1.0 A, secondary voltage of external power supply unit 12 V
Desktop power supply output .................................................................................. 12 V DC, max. 2.5 A
Electrical protection class ...................................................................................................................... II
Ingress protection rating .................................................................................................................. IP 20
Electrical safety ........................................................... in compliance with DIN EN 61010-1 (IEC 61010-1)
including CSA and UL directives
Pollution degree .................................................................................................................................... 2
Overvoltage category............................................................................................................................. II
Summary of Contents for TIRF 3