USING THE SLIDER LASER TIRF WITH CUSTOMERS' OWN LASERS
Laser TIRF 3
Laser Safety Concept
Carl Zeiss
02/2014
423683-7144-001
93
6.1
Laser Safety Concept
The Laser TIRF imaging system is equipped with interlocks on the safety-relevant laser components. The
interlocks close the laser shutter immediately if one of these components is no longer in a safe condition
or is no longer in the correct work position:
−
Slider Laser TIRF
−
Transmitted light illuminator mount
−
Eyepiece shutter
−
Reflector turret
−
Laser safety cover for reflected light
−
Camera input on sideport
−
Reflected light illumination connector for HXP 120
−
Reflected light illumination connector (optional)
−
Door interlock (laser workroom)
Legend
LS:
Laser safety
To ensure that the laser system is safe, at least one of the components in this box must be selected in line with the system
configuration. Multiple selection possible with 60N-C 1" 1.0x camera adapter.
Fig. 6-1
Laser safety concept (schematic)