USING THE SLIDER LASER TIRF WITH CUSTOMERS' OWN LASERS
Carl Zeiss
Basic kit laser safety Rev. 2
Laser TIRF 3
100
423683-7144-001
02/2014
•
Connect the 25-pin sub-D connector (Fig. 6-7/
5
) of the distribution box for laser safety (Fig. 6-7/
6
) to
the jack for Laser TIRF on the rear of the microscope stand (Fig. 6-7/
10
).
•
Connect the sub-D connector plugs to the four 15-pin sub-D ports on the distribution box (Fig. 6-7/
6
)
as follows:
−
Laser safety cover or laser safety incubator
→
Type B
jack (Fig. 6-7/
1
),
−
Laser shutter for customer's own laser module
→
Type B
connector (Fig. 6-7/
2
),
−
Slider Laser TIRF
→
Type A
connector(Fig. 6-7/
3
),
−
Laser manipulation slider
→
Type A
connector(Fig. 6-7/
3
),
−
Remote control switch
→
Type A
jack (Fig. 6-7/
4
).
•
Insert the plug (Fig. 6-7/
7
) again into the distribution box for laser safety.
The red incubators for wavelengths > 561 nm provide no protection, meaning that users must
undertake laser protection measures themselves.