8. Troubleshooting
8.1. Overview
The following table gives some clues to solve problems.
If you cannot solve the problem or if you are feeling unsure do not hesitate to get in contact with
your local Carl Zeiss service engineer.
DANGER
Danger to life: Hazardous voltage inside the workstation .
Only service engineers trained and authorised by Carl Zeiss are allowed to service the
workstation and to perform work on the electrical system of the workstation.
Keyword
Symptom
Possible reason
Recommended action/s
Vacuum
„Vac ready = OK“ is not shown
after specimen exchange.
„Vac ready = OK“ is shown very
late after specimen exchange.
System vacuum is bad due to a
vacuum leak at the chamber
door.
Gas ballast at rotary pump or
scroll pump is activated.
Check the chamber door seal for
cleanliness.
If required, replace the chamber
door seal.
Deactivate gas ballast at the pre-
vacuum pump.
Vacuum
workstation does not vent.
No nitrogen.
No compressed air.
Check nitrogen supply.
Check compressed air supply.
System
vacuum
„Vac ready = OK“ is indicated
abnormally fast.
Penning gauge has not been
identified correctly.
Restart the workstation:
If this does not solve the problem,
call your local Carl Zeiss service
engineer.
System
vacuum
Bad system vacuum.
Chamber door seal does not
seal tightly.
Check chamber door seal for par-
ticles.
Replace the chamber door seal.
Refer to section 8.2.2.
Gun vacuum
Gun vacuum is worse than 8 to
9 x 10
-9
mbar.
Gun has been switched off for
safety reasons since gun
vacuum is too bad.
Bakeout the gun head.
Refer to section 8.3.1.
Specimen
stage
Stage does not move.
Stage needs to be initialised.
Initialise the stage. Refer to sec-
tion 8.2.1.
If this does not solve the problem,
call your local Carl Zeiss service
engineer.
Specimen
stage
Stored position cannot be
approached correctly.
Stage needs to be driven to a
well-defined position.
Initialise the stage.
Refer to section 8.2.1.
Specimen
stage/
PC
Stored position cannot be
approached correctly.
PC has crashed.
Stage needs to be driven to a
well-defined position.
Initialise the stage.
Refer to section 8.2.1.
Summary of Contents for AURIGA Compact Crossbeam
Page 1: ...AURIGA Compact Crossbeam workstation Instruction Manual ...
Page 50: ...RM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP PK aÉëÅêáéíáçå ìëíçãÉê ëÉêîáÅÉ ...
Page 54: ...RQ çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP RK fåëí ää íáçå ...
Page 150: ...NRM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 156: ...NRS çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NNK ÄÄêÉîá íáçåë ...
Page 160: ...NSM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...