8.4. Power circuit
8.4.1. Checking the circuit breakers
DANGER
Danger to life: Hazardous voltage inside the workstation.
Checking and changing fuses may only be carried out by an electrically skilled person in
accordance with national safety regulations.
Turn off and lock out the workstation before opening the protective cover.
Reasons
•
Circuit breaker is tripped (lower position)
Procedure
1
Switch off the workstation completely.
a
Turn the ON/OFFswitch to OFF position.
b
Unplug the power cord.
2
Check the miniature circuit breaker (
1
).
F18 (main fuse): 16 A, AIC 10,000 rms, ON/OFF switch
3
Remove the screws (
1
).
1