3.6. Installation requirements
Location requirements
Installation site
Exclusively inside buildings
Room size
Min 3.5 m x 5.0 m x 2.3 m (W x D x H)
Service area
Min 1 m at each side
Installation category
II
Exhaust line
An exhaust line is required to remove the
wate gas of the pre-vacuum pump and to
transmit it to the outside.
Electrical supplies
Nominal AC voltage
208 V - 230 V (±10%), 1/N (L2) / PE
Protection class
I
Nominal frequency
50 - 60 Hz
Power consumption
Max. 3.5 kVA, dependent on upgrade stage
and installed optional accessories
Current input
Max. 16 A
Circuit breaker
25 A, switch off behaviour K on installation site
A special emergency-off circuit is available (part no. 340002-0167).
The EMO box is mandatory for the upgrade of FIB and/or GIS.
The EMO box should be mounted close to the system. Wiring that runs
outside the workstation should be laid and protected in appropriate cable
ducts or trays.
Moreover, it is recommended to protect the complete room by installing a
power switch next to the door of the room.
Ampere interrupting
capacity AIC
With EMO box
At least 10000 A rms
Protective ground
The workstation must be connected via a separate protective ground.
An exclusive grounding connection to earth must be provided, i. e. the
grounding terminal must not be common to other electrical equipment.
A grounding wire AWG10 (5 m long) is delivered with the workstation.
Cross section
> 4 mm
2
Ground resistance
< 0.1
Summary of Contents for AURIGA Compact Crossbeam
Page 1: ...AURIGA Compact Crossbeam workstation Instruction Manual ...
Page 50: ...RM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP PK aÉëÅêáéíáçå ìëíçãÉê ëÉêîáÅÉ ...
Page 54: ...RQ çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP RK fåëí ää íáçå ...
Page 150: ...NRM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 156: ...NRS çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NNK ÄÄêÉîá íáçåë ...
Page 160: ...NSM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...