IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times.
Always wear lint-free gloves when touching specimen, sample holder or stage. Keep the
chamber door open as short as possible.
All sample holders are equipped with a dovetail so
that the position of the sample holder is exactly de-
fined.
7
Mount the sample holder:
a
Ensure that you place the dovetail in the
correct orientation onto the holding device
on the specimen stage.
b
Make sure that the flat side of the dovetail
of the sample holder is flush with the milled
edge of the stage.
8
Look into the specimen chamber to ensure that
the specimen cannot hit any components
when it is introduced into the specimen cham-
ber.
CAUTION
Pinch hazard when closing the chamber door.
Ensure not to get your fingers caught in the chamber door gap.
9
Carefully close the chamber door.
Summary of Contents for AURIGA Compact Crossbeam
Page 1: ...AURIGA Compact Crossbeam workstation Instruction Manual ...
Page 50: ...RM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP PK aÉëÅêáéíáçå ìëíçãÉê ëÉêîáÅÉ ...
Page 54: ...RQ çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP RK fåëí ää íáçå ...
Page 150: ...NRM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 156: ...NRS çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NNK ÄÄêÉîá íáçåë ...
Page 160: ...NSM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...