IMPORTANT
The defocus value is different for each current.
b
Enter a defocus value, e.g.
20V
in the
Mill Defocus
window
4
Activate the
Force Defocus
checkbox.
The Defocus value is now applied when
using spot mode.
5
Burn a spot into the specimen for about
5 s
to
evaluate the spot quality:
Use the
Spot
mode.
6
Check the halo.
7
Slowly increase the defocus value in
10 V
steps.