A small scan frame is shown.
The image outside the scan frame is frozen. Size and position of the scan frame can be
changed by dragging and dropping.
5
Focus the image in the reduced raster.
6
Align the aperture:
a
In the
Apertures
tab, activate the
Focus
Wobble
checkbox.
The Focus Wobble is a function that sweeps the
focus of the objective lens backwards and forwards
through the focus on the specimen plane. If the
aperture is slightly misaligned, a lateral shift can be
observed.
Intensity of wobble can be adjusted by using the
Wobble Amplitude
scroll bar.
Wobble speed can be accelerated by activating
the
Wobble Fast
checkbox.
b
Click
Aperture Align
.
Use the left and right slider of the
Aperture
Align
box until there is no movement of the
detail in X- and Y- direction.
The specimen detail should just be
pulsating without shifting.
c
Deactivate the
Focus Wobble
checkbox.
Summary of Contents for AURIGA Compact Crossbeam
Page 1: ...AURIGA Compact Crossbeam workstation Instruction Manual ...
Page 50: ...RM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP PK aÉëÅêáéíáçå ìëíçãÉê ëÉêîáÅÉ ...
Page 54: ...RQ çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP RK fåëí ää íáçå ...
Page 150: ...NRM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 156: ...NRS çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NNK ÄÄêÉîá íáçåë ...
Page 160: ...NSM çÑ NSQ fåëíêìÅíáçå j åì ä ìêáÖ çãé Åí ÉåMP NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...