10 Alarms and error messages
10.2.2 Medium-priority alarm (yellow)
Supply pressure > 6 bar
Pressure of compressed gas too
high
Use compressed gas source
< 6 bar.
Switch off the device
and disconnect it from
compressed gas source.
Alarm
Cause
Remedy
The following alarms are only emitted once the respective
condition is satisfied in two consequent breathing cycles.
•
Airway pressure
/Airway pressure
•
PEEP
•
MVe
/MVe
(only with Flow measu ASB option)
•
f
(only with Flow measu ASB option)
Alarm
Cause
Remedy
Apnea mode active
No inspiration since the set apnea
alarm time
•
Check the condition of
the patient.
•
Select mandatory
ventilation mode.
Battery defective
Battery defective
Let the device run on battery
power without line power until it
switches off. Fully recharge battery
. If the device
continues to display the alarm:
Replace battery
Battery weak
Low battery status
Replace battery
Connect device to the line power
and charge battery