8 Hygienic reprocessing
2. Place the parts in a washer and disinfector.
When doing so, note:
•
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
•
Connect the hoses/tubes with the washer and disinfector.
•
All parts and lumens must be completely flushed.
•
The water must be able to run away.
3. Add cleaning agents as per the instructions for use for the
washer and disinfector.
4. If necessary: Add neutralizing agent as per the instructions for
use for the washer and disinfector.
5. Start machine reprocessing cycle according to the following
process, e.g., cycle 12 of the washer and disinfector
G 7836 CD from Miele.
6. Check the parts for residue and any remaining dirt.
7. If visible dirt remains: Repeat the machine cleaning and
disinfection.
Result
The parts are machine cleaned and disinfected.
Step
Process
Temperature
Exposure
time
Cleaning
agent/
disinfectant
1
Initial cleaning
Cold water
1 min
–
2
Main cleaning
55 °C
10 min
neodisher
®
MediClean forte
(Dr. Weigert):
0.5 %
3
Neutralization
Cold water
2 min
–
4
Rinsing
Cold water
3 min
–
5
Thermal
disinfection
90 °C
5 min
Demineralized
water
6
Drying
100 °C
25 min
–
Summary of Contents for MEDUMAT Standard2