8 Hygienic reprocessing
8.1 General instructions
•
The service life of the components of the reusable breathing
circuit is at least 30 reprocessing cycles (exception FlowCheck
sensor: typically 50 reprocessing cycles).
•
The components and accessories are not sterile on delivery.
Risk of injury and damage to property due to residues of
disinfectants and cleaning agents in the device or breathing
circuit.
Disinfectant or cleaning agent residues can get into the patient's
lung. This can result in severe or life-threatening injury to the
patient and cause material damage to the device.
All parts of the breathing circuit must be rinsed thoroughly with
water and allowed to dry completely following the hygienic
reprocessing.
After hygienic reprocessing, check the device and breathing
circuit visually for residues of cleaning agents or disinfectants
and, if necessary, remove any residues.
Perform a full check after every hygienic reprocessing.
Do not immerse the device in liquids.
Only reprocess the device hygienically with the hygiene filter
inserted.
Only clean/disinfect the filter compartment when changing
the filter.
Only clean/disinfect the filter compartment when moist and
not wet.
Damage to the device caused by ingress of liquids!
The device is rated IP54 (splash-proof). This only takes effect when
the battery is located in the battery compartment. Ingress of
liquids may damage the device, components and accessories.
Do not immerse the device, components or accessories in
liquids.
Clean the battery compartment carefully so that no liquids
enter the device.