7 Description of the modes
7.3.11 CCSV mode
Risk of injury due to unsecured airway!
If the CCSV mode is used, an unsecured airway can result in
gastric insufflation and cause injury to the patient.
Only use CCSV mode with an endotracheal tube or with
pressure-tight (blocked) tracheostomy cannulas.
Risk of injury if using a breathing circuit not approved for
CCSV mode!
Do not use a breathing circuit with reduced dead space or a 3 m
breathing circuit in CCSV mode, as the device will be unable to
correctly detect chest compressions. This may cause a fault or
treatment failure. This can injure the patient.
Only use a standard 2 m breathing circuit in CCSV mode.
Description
Abbreviation
CCSV
Long form
Chest Compression Synchronized
Ventilation
Type
Pressure-controlled
Requirement
•
Flow measu ASB option
is activated.
•
CCSV option is activated
Ventilation parameters
Left-hand navigation knob
Trigger
Central navigation knob
•
•
Right-hand navigation knob
•
CPR mode
•
pInsp
•
PEEP