182
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
8 Hygienic reprocessing
5. If the parts need to be wiped down acc. to the cleaning and
disinfection plan: Wipe parts down with a lint-free clean cloth
that has been moistened with cleaning solution at least twice.
When doing so, note:
•
Use a new cloth for each cleaning procedure.
•
Carefully wipe down all the surfaces.
•
All surfaces must be wetted with cleaning solution.
•
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
•
Rewipe uneven surfaces and grooves in particular.
6. If the parts need to be immersed acc. to the cleaning and
disinfection plan: Immerse parts in the cleaning solution.
When doing so, note:
•
Swirl the parts in the cleaning solution to coat all surfaces
and any cavities completely.
•
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
7. If visible dirt remains: Repeat the manual cleaning.
8. Rinse the parts immersed in the cleaning solution for at least
10 s and textiles for at least 2 min thoroughly with drinking-
water quality.
9. Wipe down the remaining parts with a damp cloth twice in
order to remove cleaning agent residue.