2 Safety
Notice
Material damage due to prolonged storage of the battery
without recharging!
Storing the battery for a prolonged period of time without
recharging can result in the rapid shutdown of and irreparable
damage to the battery.
When the battery is stored in the device without a power
connection: Charge battery every 3 months.
If the battery is not stored in the device: Charge battery with
SN < 20000 every 5 months and battery with SN
≥
20000
every 9 months.
2.3.3 How to use the breathing circuit
Warning
Risk of injury due to contaminated or infected breathing
circuit!
A breathing circuit which is contaminated or infected as a result of
hygienic reprocessing not being carried out or being carried out
incorrectly may transmit contamination or infections to the next
patient and harm them.
Do not reprocess disposable breathing circuits.
Perform the hygienic reprocessing of reusable breathing circuits
correctly
(see "8 Hygienic reprocessing", page 171)
.
Caution
Risk of injury from touching the contacts on the FlowCheck
sensor connection line/FlowCheck sensor connection line with
MEDUtrigger and the patient at the same time!
The contacts on the FlowCheck sensor connection line/FlowCheck
sensor connection line with MEDUtrigger are live. Touching the
contacts and the patient at the same time can injure the user or
the patient.
Do not touch the contacts on the FlowCheck sensor
connection line/FlowCheck sensor connection line with
MEDUtrigger and the patient at the same time.