8 Hygienic reprocessing
Protective sleeve
Not permitted
Not permitted
Not permitted
Washing at 60 °C in
industrial washing
machine in
accordance with
EN 16616.
Cleaning agent:
Derval SOLO (RKI)
(Kreussler)
Dosage: 2 ml/l.
Contact time:
10 min
Disinfectant:
Ottalin PERACET
(Kreussler)
Dosage: 2 ml/l
Contact time:
30 min
Hook and loop strap
with clip
Immerse in and
clean with
neodisher
®
MediClean forte
(Dr. Weigert).
Dosage: 10 ml/l,
Duration: 10 min
Not permitted
Immerse in
gigasept
®
FF neu
(Schülke)
Dosage: 50 ml/l
Exposure time:
15 min
Wash in industrial
washing machine in
accordance with
EN 16616.
Pre-washing:
5 min at 20 °C
Main washing:
30 min at 30 °C
Cleaning agent:
Turbo Usona/
Turbo Oxysan
(2-component
system)
Flow-Check sensor
connection line
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dosage: 10 ml/l,
Duration: treat all
surfaces at least
twice until visibly
clean
Wipe down with
Incidin™
OxyWipe S (Ecolab).
Exposure time:
5 min
Not permitted
Not permitted
FlowCheck sensor
connection line with
MEDUtrigger
MEDUtrigger with
connection line
Part
Manual cleaning
Wipe
disinfection
Immersion
disinfection
Mechanical
reprocessing
(Disinfection)