14 Appendix
14.1.9 Technical data on electromagnetic
compatibility (EMC)
Medical electrical equipment is subject to special precautions
in relation to electromagnetic compatibility (EMC). It must be
installed and put into operation in accordance with the EMC
information contained in the accompanying documentation.
Guidelines and manufacturer's declaration – electromagnetic interference
MEDUMAT Standard
2
is intended for operation in an electromagnetic environment as specified below.
The customer or the user of the MEDUMAT Standard
2
device must ensure that it is truly operated in such
an environment.
Interference
measurements
Compliance
Electromagnetic environment – guidelines
RF emissions
acc. to CISPR 11
Group 1, Class B
The RF emissions of MEDUMAT Standard
2
are very
low and it is unlikely that they will interfere with
neighboring electronic devices.
Emissions of harmonics
acc. to IEC 61000-3-2
Complies
MEDUMAT Standard
2
is suitable for use in all
premises including private residences and other such
facilities connected directly to the public power grid
which also supplies residential buildings.
Emissions of voltage
fluctuations/flickers
acc. to IEC 61000-3-3
Complies
RF emissions
acc. to RTCA DO-160 G
Section 21,
Category M
MEDUMAT Standard
2
is suitable for use in Category
M locations in aircraft due to its low RF emissions.
RF emissions
acc. to UN/ECE Rule No. 10
Annex 6, Annex 7
MEDUMAT Standard
2
is suitable for use in motor
vehicles due to its low RF emissions.
Interference emissions on
motor vehicle supply lines
acc. to ISO 7637-2
Complies
MEDUMAT Standard
2
is suitable for connection to
vehicle electrical systems due to its low RF emissions.