4 Preparation and operation
4
Preparation and operation
4.1 Mounting the device
The device is mounted on a portable unit as standard and is ready
for use. Observe the instructions for use of the portable units.
4.2 Connecting to a power supply
1. Check battery status
(see "4.3 Using the rechargeable
2. If necessary: Charge battery
(see "4.3.2 Charging the battery
3. Slide full battery into the battery compartment until it clicks
into place.
Loss of power due to combination of the device with
an incorrect power supply unit!
If you use a portable unit with the device combination
MEDUMAT Standard
2
and MEDUCORE Standard or
MEDUCORE Standard
2
, a loss of power may occur in the devices
in the event that these are used with a 50 W power supply unit.
Use only the more powerful 100 W power supply unit when
combining the devices MEDUMAT Standard
2
and
MEDUCORE Standard.