14 Appendix
Pressure drop [hPa] over the inspiratory and expiratory flow path at different flow
rates [l/min] acc. to EN 794-3 (in combination with MEDUMAT Standard
2
, measuring
point 1: Patient connection opening)
Breathing circuits (2 m) with FlowCheck sensor and with CO
2
measurement
Flow
[l/min]
Breathing circuit
(reusable), 2 m
WM 29190
Breathing circuit
(disposable), 2 m
WM 29192
Breathing
circuit
(disposable),
2 m, with
reduced dead
space
WM 29199
With
elbow
Without
elbow
With
elbow
Without
elbow
With
elbow
Spontaneous
respiration
in event of
power failure,
inspiratory
(STP)
(1)
2.5
1.25
1.03
0.52
-
0.33
15
2.64
2.45
2.10
-
1.43
30
3.77
3.39
4.29
-
3.93
Spontaneous
respiration
in event of
power failure,
expiratory
(BTPS)
(2)
2.5
0.43
0.41
0.95
-
0.88
15
1.68
1.66
1.63
-
1.94
30
2.68
2.56
2.34
-
3.40
Normal
operation,
inspiratory
(STP)
(1)
5
0.18
0.18
0.43
-
0.08
30
1.11
1.05
1.60
-
1.28
60
2.83
2.55
3.94
-
5.90
Normal
operation,
expiratory
(BTPS)
(2)
5
0.94
1.01
1.24
-
1.09
30
2.85
2.79
2.34
-
3.40
60
4.53
4.09
3.89
-
8.11