WM
6807
1
0
5
/2
021
MEDUMAT Standard
2
EN
29
3 Description
No.
Designation
Description
Pressure gauge
•
Indicates ventilation pressure progress.
•
Indicates pMax
as a dotted line.
•
Indicates the currently attuning airway pressure
as a green area.
•
Indicates the maximum airway pressure in the
middle.
•
Indicates the end-tidal CO
2
concentration (etCO
2
)
in the middle (only with Capnography option).
End-tidal CO
2
concentration (etCO
2
)
(only with Capnography option)
Indicates the end-tidal CO
2
concentration.
If the Capnography option is deactivated,
the peak pressure is shown here (pPeak).
Time
Displays the current time.
100 % O
2
Air Mix
Indicates whether operation with 100 % oxygen
or Air Mix mode is activated.
Ventilation mode indicator
Indicates the currently selected ventilation mode.
Inspiration time (Ti)
Indicates the inspiration time.
If an alarm is displayed, this information is omitted.
Peak pressure (pPeak)
Indicates the maximum pressure.
Plateau pressure (pPlat)
Indicates the pressure during the plateau time.
Mean pressure (pMean)
Indicates the mean pressure over all measurements.
Blue arrow
Provides access to the application menu
(turn or press the right-hand navigation knob).
Pressure curve (only with Flow
measu ASB option and Curve
display option or Capnography option)
Indicates the pressure progress.
Flow curve (only with Flow
measu ASB option)
Indicates the flow progress.
Leakage (Vleak) (only with Flow
measu ASB option)
Indicates leaks.
Positive end-expiratory pressure (PEEP)
•
Indicates the positive end-expiratory pressure.
•
Enables the positive end-expiratory pressure to
be set.
Frequency (Freq.)
•
Indicates the ventilation rate.
•
Enables the ventilation rate to be set.
Tidal volume (Vt)
•
Indicates the tidal volume.
•
Enables the tidal volume to be set.
Expiratory minute volume (MVe) (only
with Flow measu ASB option)
Indicates the expiratory minute volume and
the associated alarm limits.