MEDUMAT Standard
2
EN
77
4 Preparation and operation
4.7.8 Performing CO
2
monitoring (only with
Capnography option)
Requirement
•
The device is switched on.
•
CO
2
monitoring
was set as the ventilation mode for a new
patient
(see "4.7.3 Selecting a ventilation mode for a new
•
An etCO
2
/O
2
nasal cannula is connected
(see "4.4.7 Connecting the etCO
or
•
The CO
2
measuring tube is connected to an external interface
(e.g., bag-valve mask with breathing system filter).
1. Assess CO
2
measurements diagnostically.
2. If necessary: Configure etCO
2
Trend
(only with Capnography option)", page 115)
.
Result
The CO
2
measurements and the respiratory rate of the patient are
monitored.
When using an etCO
2
/O
2
nasal cannula, the CO
2
measurement
may be distorted by the intake of additional air. Only the CO
2
trend can be assessed.