10 Alarms and error messages
Device fault
Temporary device fault
•
Switch device off
and back on again
•
Perform a function check.
Device defective
Have the device repaired.
Device temperature
Device temperature < -20 °C
Operate device within
permitted temperature range
.
Device temperature critical
Device temperature > 75 °C
Operate device within
permitted temperature range
.
Leakage on patient side
(only with CCSV option)
Leakage on the breathing circuit or
endotracheal tube
Check the breathing circuit
and endotracheal tube.
MVe
(only with Flow
measu ASB option)
Upper limit value exceeded
•
Check the condition of
the patient.
•
Check the set limit values
for plausibility.
MVe
(only with Flow
measu ASB option)
Lower limit value not reached
PEEP
Obstruction of the patient
'
s
airways
Free the patient
'
s airways of
obstructions.
Endotracheal tube wrongly
positioned
Position endotracheal tube
correctly.
Hoses/tubes kinked or pinched
Route hoses/tubes so that they
are not kinked or pinched.
Patient valve defective
Replace patient valve.
Ventilation settings incorrectly set Adjust ventilation settings.
Reconnect patient to device
The device self-test is not
complete.
•
Disconnect the device from
the patient.
•
Switch device off
and back on again
•
Reconnect patient to device.
Supply pressure < 2.7 bar
Oxygen cylinder not opened
Open oxygen cylinder.
Oxygen cylinder almost empty
Replace oxygen cylinder.
Compressed gas source not
connected correctly
Connect compressed gas source
correctly.
Compressed gas source defective Replace compressed gas source.
Low-pressure hose kinked or
pinched
Route the low-pressure hose so
that it is not kinked or pinched.
Pressure reducer defective
Replace pressure reducer.
Alarm
Cause
Remedy