200
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
9 Function check
9
Function check
9.1 Intervals
Carry out a function check at regular intervals:
9.2 Preparing for the function check
1. Check battery status: The battery must be fully charged.
If necessary: Charge or replace the battery.
2. Check the following parts for external damage:
•
Device
•
Plug and cable
•
Breathing circuit
•
Accessories
If necessary: Replace parts.
3. Check the patient valve of the breathing circuit
(see "9.5 Testing the reusable breathing circuit", page 210)
.
If necessary: Replace the breathing circuit.
4. Check the oxygen level in the oxygen cylinder.
If necessary: Change the oxygen cylinder.
Part concerned
Interval
Device
•
Before each use
•
After each hygienic reprocessing
•
After each repair
Breathing circuit (reusable
breathing circuit)
•
Before each use
•
After each hygienic reprocessing
•
After each disassembly
•
At least every 6 months
Summary of Contents for MEDUMAT Standard2
Page 1: ...MEDUMAT Standard2 Ventilator Instructions for Use for Devices from Software Version 5 1...
Page 269: ......
Page 270: ......
Page 271: ......