236
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
14 Appendix
(1)
BTPS (Body Temperature and Pressure, saturated): Volume at current
ambient pressure and 37 °C, with 100 % saturated gas
Trigger window for mandatory
mechanical breaths
20 % of Te (SIMV, SIMV + ASB, B ASB and PRVC + ASB)
0 % to 100 % of Te (S-IPPV)
Trigger window for ASB mechanical
breaths
0 % to 100 % Te
Trigger window aPCV
(only with Pressure controlled
ventilation modes option)
0 % to 100 % Te (adjustable)
Volume monitoring
(1)
Measurement range: 40 ml to 8000 ml
Tolerance: ± 15 %
Airway pressure monitoring
Measurement range: -5 mbar to +80 mbar
Tolerance: ± 3 mbar
Oxygen concentration:
•
Air Mix operation
•
No Air Mix operation
See "14.1.8 Oxygen concentration with Air Mix operation",
page 251.
100 % oxygen
Concentrator oxygen (90 % to 96 % oxygen)
Pressurized gas thread
External thread G 3/8
Connection for ventilation hose
WEINMANN Emergency-specific
Patient valve connections
WEINMANN Emergency-specific
Service life of hygiene filter
24 h in Air Mix operation or 6 months
Efficiency of hygiene filter
> 99 % for viruses and bacteria (particle size > 27 nm)
Specification
Device
Subject to alterations in design.