2 Safety
Risk of injury from switching on a device with activated NVG
mode during daylight or without a night vision device!
A device with activated NVG mode cannot be used straight away
during daylight or without a night vision device. This can injure the
patient.
Keep an alternative ventilation unit at the ready.
Risk of injury if CCSV mode is used on infants!
Use of the CCSV mode can result in increased intrathoracic
pressures and thus injure infants' lungs.
Do not use CCSV mode on patients weighing less than 10 kg.
Risk of injury from use of pneumatic nebulizers during
volume-controlled ventilation!
The use of pneumatic nebulizers increases the minute volume
administered to the patient.
Do not use pneumatic nebulizers during volume-controlled
ventilation.
Caution
Risk of injury due to operation of the device with compressed
air!
During operation with compressed air, the volume delivered by the
device is excessively high and the oxygen concentration of the
output is too low. This may lead to volutrauma and hypoxia in the
patient.
Only operate the device with medical oxygen or concentrator
oxygen.
Risk of injury due to drying out of the airways!
Prolonged ventilation using the device may dry out the airways of
the patient and cause them an injury.
Do not use the device for long-term ventilation.
Risk of injury due to unsuitable concentrator gas!
Unsuitable concentrator gas may distort treatment and result in
injury to the patient.
Only use concentrator oxygen (90 % to 96 % oxygen) or
medical oxygen.
Risk of injury if the patient valve is covered!
The patient valve may be covered due to the position of the
patient and prevented from functioning properly.
Always keep the patient valve clear.