7 Description of the modes
7.4 Additional functions
7.4.1 CPR mode
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
2
monitoring are output in the additional functions CPR
and RSI. This can injure the patient.
Notify the user if the alarms in the additional functions CPR
and RSI have been deactivated.
Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
Description
Abbreviation
CPR Manual CPR IPPV
CPR CCSV
Long form
Cardiopulmonary Resuscitation
Type
Volume-controlled
Pressure-controlled
Requirement
None
None
•
Flow
measu
ASB option is
activated
•
CCSV option is
activated
Ventilation parameters
Left-hand navigation
knob
Vt
Vt
Trigger
Central navigation
knob
-
Freq.
•
•
Right-hand navigation
knob
•
CPR
mode
•
pMax
•
CPR
mode
•
Pause
•
pMax
•
PEEP
•
CPR mode
•
pInsp
•
PEEP